Abstract

Background: A common clinical scenario encountered by an orthopaedic surgeon is a patient with a secure cementless acetabular shell and a failed polyethylene liner. One treatment option is to cement a new liner into the fixed shell. The purpose of this study was to evaluate technical variables to improve the mechanical strength of such cemented liner constructs.

Methods: The contributions of shell texturing, liner texturing, and cement mantle thickness (between the liner and the shell) were evaluated by comparing torsional strength (among nine groups of constructs) and lever-out strength (among eight groups of constructs).

Results: Failure almost always occurred at the cement-liner interface. The two exceptions (failure at the shell-cement interface) occurred with a polished, untextured shell with no screw-holes. This finding indicates that if a shell has existing texturing (such as holes), further intraoperative scoring of the shell is unnecessary, but some sort of texturing is necessary to avoid construct failure at the shell-cement interface. Textured liners had significantly (a = 0.05) greater torsional and lever-out strength than untextured liners. The greatest construct strength occurred when liner grooves were oriented so as to oppose the applied loading. A 4-mm-thick cement mantle resulted in slightly greater torsional strength than a 2-mm-thick cement mantle, and a 2-mm-thick cement mantle resulted in considerably greater lever-out strength than a 4-mm-thick cement mantle, but these differences were not significant.

Conclusions: When cementing a liner into a well-fixed shell, a surgeon should ensure that both the shell and the liner are textured, as interdigitation of the cement with the shell and the liner is crucial to the mechanical strength of this construct.

Citing Articles

Cementation of a Metal Dual Mobility Liner in Patients Undergoing Revision Total Hip Arthroplasty.

Chen Z, Bains S, Dubin J, Sax O, Gilson G, Jacobstein D Arthroplast Today. 2023; 24:101270.

PMID: 38077925 PMC: 10704222. DOI: 10.1016/j.artd.2023.101270.


Cementation of a dual mobility cup in a well-fixed acetabular component- a reliable option in revision total hip arthroplasty?.

Bellova P, Koch F, Stiehler M, Hartmann A, Fritzsche H, Gunther K BMC Musculoskelet Disord. 2021; 22(1):982.

PMID: 34819056 PMC: 8613958. DOI: 10.1186/s12891-021-04835-z.


The Utility of a Plain Film Arthrogram to Confirm Acute Liner Dissociation in the Setting of Primary Total Hip Arthroplasty.

McQuail P, Miller P, Nolan P, Ellanti P, McCarthy T Cureus. 2020; 12(8):e9951.

PMID: 32983657 PMC: 7510184. DOI: 10.7759/cureus.9951.


The Effects of Cyclic Loading and Motion on the Implant-Cement Interface and Cement Mantle of PEEK and Cobalt-Chromium Femoral Total Knee Arthroplasty Implants: A Preliminary Study.

de Ruiter L, Cowie R, Jennings L, Briscoe A, Janssen D, Verdonschot N Materials (Basel). 2020; 13(15).

PMID: 32722599 PMC: 7435552. DOI: 10.3390/ma13153323.


Cementing Constrained Liners Into Secure Cementless Shells: A Minimum 15-Year Follow-Up Study.

Young G, Abdel M, Amendola R, Goetz D, Lewallen D, Callaghan J J Arthroplasty. 2017; 32(11):3480-3483.

PMID: 28780226 PMC: 6322832. DOI: 10.1016/j.arth.2017.07.014.