Three Dimensional Inertial Sensing of Foot Movements for Automatic Tuning of a Two-channel Implantable Drop-foot Stimulator

Overview
Journal Med Eng Phys
Date 2002 Dec 18
PMID 12485783
Citations 14
Authors
Affiliations
Soon will be listed here.
Abstract

A three dimensional inertial sensing system for measuring foot movements during gait is proposed and tested. It can form the basis for an automated tuning system for a two-channel implantable drop-foot stimulator. The foot orientation and position during the swing phase of gait can be reconstructed on the basis of three-dimensional measurement of acceleration and angular velocity, using initial and final conditions during mid-stance. The foot movements during gait of one stroke person using the implanted two-channel stimulator were evaluated for several combinations of stimulation parameters for both channels. The reconstructed foot movements during gait in this person indicated that the channel stimulating the deep peroneal nerve contributes mainly to dorsiflexion and provides some reduction of inversion seen without stimulation, while the channel activating the superficial peroneal nerve mainly provides additional reduction of inversion. This agrees with anatomical knowledge about the function of the muscles activated by both branches of the peroneal nerve. The inertial sensor method is expected to be useful for the clinical evaluation of foot movements during gait supported by the two-channel drop-foot stimulator. Furthermore, it is expected to be applicable for the automated balancing of the two stimulation channels to ensure optimal support of gait.

Citing Articles

Can inertial measurement unit sensors evaluate foot kinematics in drop foot patients using functional electrical stimulation?.

dAndrea F, Taylor P, Yang K, Heller B Front Hum Neurosci. 2023; 17:1225086.

PMID: 38021225 PMC: 10666752. DOI: 10.3389/fnhum.2023.1225086.


Machine learning aided jump height estimate democratization through smartphone measures.

Mascia G, De Lazzari B, Camomilla V Front Sports Act Living. 2023; 5:1112739.

PMID: 36845828 PMC: 9947475. DOI: 10.3389/fspor.2023.1112739.


Assessment of shoulder range of motion using a commercially available wearable sensor-a validation study.

Chan L, Chua C, Chou S, Seah R, Huang Y, Luo Y Mhealth. 2022; 8:30.

PMID: 36338310 PMC: 9634209. DOI: 10.21037/mhealth-22-7.


Estimation of Foot Trajectory and Stride Length during Level Ground Running Using Foot-Mounted Inertial Measurement Units.

Suzuki Y, Hahn M, Enomoto Y Sensors (Basel). 2022; 22(19).

PMID: 36236228 PMC: 9573471. DOI: 10.3390/s22197129.


Motor neuroprosthesis for promoting recovery of function after stroke.

Mendes L, Lima I, Souza T, do Nascimento G, Resqueti V, Fregonezi G Cochrane Database Syst Rev. 2020; 1:CD012991.

PMID: 31935047 PMC: 6984639. DOI: 10.1002/14651858.CD012991.pub2.