Abstract

Purpose: Evidence is emerging that the abrupt compliance mismatch that exists at the junction between the stent ends and the host arterial wall disturbs both the vascular hemodynamics and the natural wall stress distribution. These stent-induced alterations are greatly reduced by smoothing the compliance mismatch between the stent and host vessel. A stent that provides this smooth transition in compliance, the compliance matching stent (CMS), has been developed. This study attempts to evaluate the hemodynamics and wall mechanical consequences of the CMS both in vitro and in vivo.

Materials And Methods: Finite element analysis was used to assess the solid mechanical behavior (compliance and stress) of the CMS in a stent/artery hybrid structure. A similar analysis was performed with a Palmaz stent. In vivo hemodynamics and wall mechanical changes induced by the CMS were investigated in a swine model from direct measurements of flow, pressure, diameter, and histology in the stented segment of superficial femoral arteries after 7 days.

Results: Finite element analysis showed that the abrupt compliance mismatch was substantially smoothed between the vessel portions with and without the stent with CMS segments. Circumferential stress was also markedly reduced with the CMS compared to other stent. The in vivo results showed that the CMS was efficient in compliance matching and did not dampen flow or pressure waves downstream the stent. Concurrent histology showed limited thrombus and inflammatory cell accumulation around the stent struts.

Conclusion: These results indicate that the stent/artery hybrid structure can be compliance matched with proper stent design and that this structure limits solid mechanical stress and hemodynamic disturbances. It remains to be seen whether compliance-matched vascular stents reduce in-stent restenosis.

Citing Articles

Manufacturing, Processing, and Characterization of Self-Expanding Metallic Stents: A Comprehensive Review.

Vanaei S, Hashemi M, Solouk A, Ilani M, Amili O, Hefzy M Bioengineering (Basel). 2024; 11(10).

PMID: 39451359 PMC: 11505524. DOI: 10.3390/bioengineering11100983.


The Ability to Look Beyond: The Treatment of Peripheral Arterial Disease.

Sirignano P, Margheritini C, Ruggiero F, Panzano C, Filippi F, Rizzo L J Clin Med. 2023; 12(9).

PMID: 37176513 PMC: 10179057. DOI: 10.3390/jcm12093073.


Numerical Analysis for Non-Uniformity of Balloon-Expandable Stent Deployment Driven by Dogboning and Foreshortening.

Rahinj G, Chauhan H, Sirivella M, Satyanarayana M, Ramanan L Cardiovasc Eng Technol. 2021; 13(2):247-264.

PMID: 34431035 DOI: 10.1007/s13239-021-00573-4.


Mechanical and physio-biological properties of peptide-coated stent for re-endothelialization.

Bae I, Jeong M, Park D, Lim K, Shim J, Kim M Biomater Res. 2020; 24:4.

PMID: 31998531 PMC: 6979279. DOI: 10.1186/s40824-020-0182-x.


Cross-sectional pinching in human femoropopliteal arteries due to limb flexion, and stent design optimization for maximum cross-sectional opening and minimum intramural stresses.

Desyatova A, Poulson W, MacTaggart J, Maleckis K, Kamenskiy A J R Soc Interface. 2018; 15(145).

PMID: 30135264 PMC: 6127163. DOI: 10.1098/rsif.2018.0475.