Using a Sterile Disposable Protective Surgical Drape for Reduction of Radiation Exposure to Interventionalists

Overview
Specialties Oncology
Radiology
Date 2002 Jan 5
PMID 11756110
Citations 18
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: The purpose of this paper is to show the effectiveness of a new radiation protection method designed to decrease the amount of scatter radiation received by practitioners performing procedures under fluoroscopic guidance.

Materials And Methods: A sterile, disposable, lead-free surgical drape containing radiation protection material composed primarily of bismuth was evaluated for effectiveness in reducing radiation doses to health care personnel. Measurements of phantom scatter, patient scatter, skin entrance, and the effects of collimation, together with comparative monthly thermoluminescent dosimeter recordings, were taken to determine the effectiveness of X-ray beam attenuation using the bismuth drapes.

Results: Scatter radiation to physicians, as measured by thermoluminescent dosimeters placed on each eye, the thyroid, and the wrist, was reduced by 12-fold for the eyes, 25-fold for the thyroid, and 29-fold for the hands when the radiation-attenuating surgical drape was used when compared with control studies performed with a standard nonattenuating surgical drape alone. Monthly thermoluminescent dosimeter measurements decreased fourfold in one physician. Using the protective drape reduced exposure to the assistant in each case to negligible levels. Skin entrance dose was not increased unless the protective drape was placed directly in the X-ray beam. An X-ray attenuation factor equivalent to 0.1 mm of lead with 8 x 8 cm collimation reduced the scatter rates from five- to ninefold despite a 30-40% increase in entrance exposure rate as the lead equivalence increased.

Conclusion: Depending on the procedure, the height of the practitioner, and the positioning of the radiation-attenuating surgical drape, use of this drape can substantially reduce the radiation dose to personnel with minimal or no additional radiation exposure to the patient.

Citing Articles

Addressing the Occupational Risk of Radiation Exposure in the Evolving Field of Interventional Echocardiography.

Lauten P, Lapp H, Goebel B Struct Heart. 2024; 8(5):100328.

PMID: 39290673 PMC: 11403041. DOI: 10.1016/j.shj.2024.100328.


PICS/AEPC/APPCS/CSANZ/SCAI/SOLACI: Expert Consensus Statement on Cardiac Catheterization for Pediatric Patients and Adults With Congenital Heart Disease.

Holzer R, Bergersen L, Thomson J, Aboulhosn J, Aggarwal V, Akagi T J Soc Cardiovasc Angiogr Interv. 2024; 3(1):101181.

PMID: 39131968 PMC: 11307799. DOI: 10.1016/j.jscai.2023.101181.


Assessing the Efficacy of RADPAD Protection Drape in Reducing Radiation Exposure to Operators in the Cardiac Catheterization Laboratory: A Systematic Review and Meta-Analysis.

Bahar A, Khanal R, Hamza M, Goru R, Shafiq A, Haider M Cureus. 2024; 16(4):e59215.

PMID: 38807800 PMC: 11132176. DOI: 10.7759/cureus.59215.


Minimizing scattered radiation dose in cardiac catheterization laboratory during interventional procedures using lead free drape - MILD study.

Bhat K, Guleria V, Singla M, Bohra V, Kumar J R, Bharadwaj P Indian Heart J. 2022; 74(3):201-205.

PMID: 35427629 PMC: 9243615. DOI: 10.1016/j.ihj.2022.04.001.


Radiation Protection in Interventional Radiology.

Garg T, Shrigiriwar A Indian J Radiol Imaging. 2022; 31(4):939-945.

PMID: 35136507 PMC: 8817818. DOI: 10.1055/s-0041-1741049.