Abstract

We developed a system for amperometric detection of Escherichia coli (E. coli) based on the integration of microelectromechanical systems (MEMS), self-assembled monolayers (SAMS), DNA hybridization, and enzyme amplification. Using MEMS technology, a detector array was fabricated which has multiple electrodes deposited on a Si wafer and was fully reusable. Using SAMs, a monolayer of the protein streptavidin was immobilized on the working electrode (Au) surface to capture rRNA from E. coli. Three different approaches can be used to immobilize streptavidin onto Au, direct adsorption of the protein on bare Au, binding the protein to a biotinylated thiol SAM on Au, and binding the protein to a biotinylated disulfide monolayer on Au. The biotinylated thiol approach yielded the best results. High specificity for E. coli was achieved using ssDNA-rRNA hybridization and high sensitivity was achieved using enzymatic amplification with peroxidase as the enzyme. The analysis protocol can be conducted with solution volumes on the order of a few microliters and completed in 40 min. The detection system was capable of detecting 1000 E. coli cells without polymerase chain reaction with high specificity for E. coli vs. the bacteria Bordetella bronchiseptica.

Citing Articles

Selective Quantification of Bacteria in Mixtures by Using Glycosylated Polypyrrole/Hydrogel Nanolayers.

Balser S, Rohrl M, Spormann C, Lindhorst T, Terfort A ACS Appl Mater Interfaces. 2024; 16(11):14243-14251.

PMID: 38442898 PMC: 10959108. DOI: 10.1021/acsami.3c14387.


Waterborne pathogens detection technologies: advances, challenges, and future perspectives.

Oon Y, Oon Y, Ayaz M, Deng M, Li L, Song K Front Microbiol. 2023; 14:1286923.

PMID: 38075917 PMC: 10708915. DOI: 10.3389/fmicb.2023.1286923.


Direct-from-specimen microbial growth inhibition spectrums under antibiotic exposure and comparison to conventional antimicrobial susceptibility testing.

Chen J, San S, Kung A, Tomasek M, Liu D, Rodgers W PLoS One. 2022; 17(2):e0263868.

PMID: 35171945 PMC: 8849476. DOI: 10.1371/journal.pone.0263868.


Rapid Electrochemical-Based PCR-Less Microbial Quantification and Antimicrobial Susceptibility Profiling Directly From Blood and Urine With Unknown Microbial Load or Species.

Chen J, Navarro E, Nunez E, Gau V Front Bioeng Biotechnol. 2021; 9:744198.

PMID: 34604191 PMC: 8481646. DOI: 10.3389/fbioe.2021.744198.


Interdigitated Electrode Biosensor Based on Plasma-Deposited TiO Nanoparticles for Detecting DNA.

Yoo J, Jeong H, Park S, Park S, Lee J Biosensors (Basel). 2021; 11(7).

PMID: 34209744 PMC: 8301939. DOI: 10.3390/bios11070212.