US-guided Nephrostomy with the Aid of a Magnetic Field-based Navigation Device in the Porcine Pelvicaliceal System

Overview
Date 2001 May 8
PMID 11340143
Citations 6
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To assess the technique and feasibility of ultrasound (US)-guided nephrostomy performed with the aid of a magnetic field-based navigation device.

Materials And Methods: Twelve percutaneous nephrostomy procedures were performed in seven domestic pigs under real-time US guidance employing a navigation system. Two magnetic field sensors registered the actual position of the puncture needle and the US transducer. The actual needle position and the aimed puncture tract were graphically projected into the real-time US images, allowing exact navigation of the needle. After placement of the needle into the renal pelvis, the nephrostomy tube was inserted by means of the Seldinger technique.

Results: The puncture procedure was successfully performed in each case. The renal pelvis was punctured in a single attempt in 10 of 12 cases. In two cases, multiple attempts were necessary because bending of the puncture needle led to inaccuracies of the navigation system. With exception of two small perirenal urinomas, no complications were seen.

Conclusions: The magnetic field-based navigation device allows easy and accurate performance of nephrostomy guided solely by US.

Citing Articles

Deep-learning-aided forward optical coherence tomography endoscope for percutaneous nephrostomy guidance.

Wang C, Calle P, Tran Ton N, Zhang Z, Yan F, Donaldson A Biomed Opt Express. 2021; 12(4):2404-2418.

PMID: 33996237 PMC: 8086467. DOI: 10.1364/BOE.421299.


AngleNav: MEMS Tracker to Facilitate CT-Guided Puncture.

Li R, Xu S, Pritchard W, Karanian J, Krishnasamy V, Wood B Ann Biomed Eng. 2018; 46(3):452-463.

PMID: 29305735 PMC: 7641877. DOI: 10.1007/s10439-017-1968-4.


Percutaneous puncture of renal calyxes guided by a novel device coupled with ultrasound.

Chan C, Srougi V, Tanno F, Jordao R, Srougi M Int Braz J Urol. 2015; 41(5):953-8.

PMID: 26689521 PMC: 4756972. DOI: 10.1590/S1677-5538.IBJU.2014.0586.


[Placement of percutaneous nephrostomy by open magnetic resonance imaging: clinical results and current status in urology].

Porsch M, Wendler J, Fischbach F, Schindele D, Janitzky A, Baumunk D Urologe A. 2012; 51(12):1722-7.

PMID: 23086482 DOI: 10.1007/s00120-012-3035-y.


Robot arm based flat panel CT-guided electromagnetic tracked spine interventions: phantom and animal model experiments.

Penzkofer T, Isfort P, Bruners P, Wiemann C, Kyriakou Y, Kalender W Eur Radiol. 2010; 20(11):2656-62.

PMID: 20567831 DOI: 10.1007/s00330-010-1837-0.