» Articles » PMID: 10916262

Chronic Neural Stimulation with Thin-film, Iridium Oxide Electrodes

Overview
Date 2000 Aug 1
PMID 10916262
Citations 45
Authors
Affiliations
Soon will be listed here.
Abstract

Experiments were conducted to assess the effect of chronic stimulation on the electrical properties of the electrode-tissue system, as measured using electrochemical impedance spectroscopy (EIS) and cyclic voltammetry (CV). Silicon, micromachined probes with multiple iridium oxide stimulating electrodes (400-1600 micron 2) were implanted in guinea pig cortex. A 10-17 day post-operative recovery period was followed by five days of monopolar stimulation, two hours/electrode each day using biphasic, constant current stimulation (5-100 microA, 100 microseconds/phase). EIS and CV data were taken before and after stimulation. The post-stimulation impedance [at mid-range frequencies (100 Hz-100 kHz)] consistently and significantly decreased relative to prestimulation levels. Impedance magnitude increased permanently at low frequencies (< 100 Hz), correlating to a change in the charge storage capacity (the area under a cyclic voltammagram). Impedance magnitude significantly increased during the recovery period, though this increase could be mostly reversed by applying small currents. A mathematical model of the electrode-tissue system impedance was used to analyze in vivo behavior. The data and modeling results shows that applying charge to the electrode can consistently reduce the impedance of the electrode-tissue system. Analysis of explanted probes suggests that the interaction between the tissue and electrode is dependent on whether chronic pulses were applied. It is hypothesized that the interface between the tissue and metal is altered by current pulsing, resulting in a temporary impedance shift.

Citing Articles

Thin-film implants for bioelectronic medicine.

Oldroyd P, El Hadwe S, Barone D, Malliaras G MRS Bull. 2024; 49(10):1045-1058.

PMID: 39397879 PMC: 11469980. DOI: 10.1557/s43577-024-00786-7.


Electrical rejuvenation of chronically implanted macroelectrodes in nonhuman primates.

OSullivan K, Orazem M, Otto K, Butson C, Baker J J Neural Eng. 2024; 21(3).

PMID: 38862007 PMC: 11302379. DOI: 10.1088/1741-2552/ad5703.


Bioelectronic Neural Interfaces: Improving Neuromodulation Through Organic Conductive Coatings.

Duan W, Aregueta Robles U, Poole-Warren L, Esrafilzadeh D Adv Sci (Weinh). 2023; 11(27):e2306275.

PMID: 38115740 PMC: 11251570. DOI: 10.1002/advs.202306275.


Generation of direct current electrical fields as regenerative therapy for spinal cord injury: A review.

Matter L, Harland B, Raos B, Svirskis D, Asplund M APL Bioeng. 2023; 7(3):031505.

PMID: 37736015 PMC: 10511262. DOI: 10.1063/5.0152669.


Translational opportunities and challenges of invasive electrodes for neural interfaces.

Shen K, Chen O, Edmunds J, Piech D, Maharbiz M Nat Biomed Eng. 2023; 7(4):424-442.

PMID: 37081142 DOI: 10.1038/s41551-023-01021-5.