Abstract

Background: Signal-averaged ECG (SAECG) reproducibility is reported to have a component that is independent of residual noise. Methods and Results-In group 1, multiple paired SAECGs were obtained to noise levels of 0.3+/-0.1 and 0.5+/-0.2 microV. For the 0.5- and 0. 3-microV noise recordings, QRS duration (QRSd) was 101.2+/-11.3 and 104.6+/-9.6 ms, respectively (P<0.0001), and the differences in paired QRSd (DeltaQRSd) were normally distributed, with variances of 11.4 and 26.2 ms(2) (P<0.0001). Paired SAECGs were obtained in group 2 patients without and with late potentials; DeltaQRSd variance was 3.3 and 217.9 ms(2) (P<0.0001). In group 3, >/=10 SAECGs were acquired at noise levels of 0.2 to 0.8 microV, in 0.1-microV increments. QRSd increased as noise level decreased. The variance was greater in low-noise (0.2 to 0.4 microV) versus higher-noise (0. 5 to 0.8 microV) recordings. In group 4, SAECGs were analyzed with bidirectional and Bispec filters, with no difference in QRSd between the 2 filters and a normally distributed DeltaQRSd. A computer simulation demonstrated that alterations in the phase relationship of noise to signal results in a normal distribution of signal end points.

Conclusions: Within the acceptable noise range for SAECG, lower noise results in longer QRSd and larger variance, suggesting that more accurate recordings may have less reproducibility. The random timing of noise relative to signal results in the distribution/variance of repeated measurements. Statistical strategies may be used to reduce some of this variance and may enhance the diagnostic utility of SAECG.

Citing Articles

Classification of Parkinson's disease severity using gait stance signals in a spatiotemporal deep learning classifier.

Munoz-Mata B, Dorantes-Mendez G, Pina-Ramirez O Med Biol Eng Comput. 2024; 62(11):3493-3506.

PMID: 38884852 DOI: 10.1007/s11517-024-03148-2.


Prior Mental Fatigue Impairs Marksmanship Decision Performance.

Head J, Tenan M, Tweedell A, LaFiandra M, Morelli F, Wilson K Front Physiol. 2017; 8:680.

PMID: 28951724 PMC: 5599781. DOI: 10.3389/fphys.2017.00680.


Feasibility study of transfer function model on electrocardiogram change caused by acupuncture.

Lee H, Kim H, Kim J, Oh H, Park Y, Park Y BMC Complement Altern Med. 2017; 17(1):101.

PMID: 28178964 PMC: 5299648. DOI: 10.1186/s12906-017-1615-5.


Improving clustering by imposing network information.

Gerber S, Horenko I Sci Adv. 2015; 1(7):e1500163.

PMID: 26601225 PMC: 4643807. DOI: 10.1126/sciadv.1500163.


Low noise level unmasks late potentials on signal-averaged electrocardiography.

Frances R Exp Clin Cardiol. 2010; 15(3):e61-4.

PMID: 20959882 PMC: 2954031.