» Articles » PMID: 10864608

Camino Intracranial Pressure Monitor: Prospective Study of Accuracy and Complications

Overview
Date 2000 Jun 23
PMID 10864608
Citations 20
Authors
Affiliations
Soon will be listed here.
Abstract

Objectives: The fibreoptic device is a type of intracranial pressure monitor which seems to offer certain advantages over conventional monitoring systems. This study was undertaken to analyse the accuracy, drift characteristics, and complications of the Camino fibreoptic device.

Methods: One hundred and eight Camino intracranial pressure (ICP) devices, in their three modalities, were implanted during 1997. The most frequent indication for monitoring was severe head injury due to road traffic accidents.

Results: Sixty eight probe tips were cultured; 13.2% of the cases had a positive culture without clinical signs of infection, and 2.9% had a positive culture with clinical signs of ventriculitis. The most common isolated pathogen was Staphylococcus epidermidis. All patients were under cephalosporin prophylaxis during monitoring. Haemorrhage rate in patients without coagulation disorders was 2.1% and 15.3% in patients with coagulation abnormalities. Drift characteristics were studied in 56 cases; there was no drifting from the values expected according to the manufacturer's specifications in 34 probes. There was no relation between direction of the drift and duration of placement, nor between drift and time.

Conclusions: Although the complication and drift rates were similar to those reported elsewhere, there was no correlation between the direction of the drift and long term monitoring despite the fact that some published papers refer to overestimation of values with time with this type of device.

Citing Articles

Efficacy and safety of bedside percutaneous three-millimeter twist-drill trephination under local anesthesia-a retrospective study of 1000 patients.

Puhahn-Schmeiser B, Wegent H, Won S, Zentner J, Freiman T Acta Neurochir (Wien). 2024; 166(1):87.

PMID: 38366108 DOI: 10.1007/s00701-024-05958-7.


Ultrasound detected increase in optic disk height to identify elevated intracranial pressure: a systematic review.

Ghanem G, Haase D, Brzezinski A, Ogawa R, Asachi P, Chiem A Ultrasound J. 2023; 15(1):26.

PMID: 37227512 PMC: 10212868. DOI: 10.1186/s13089-023-00324-7.


Intracranial pressure monitoring in neurosurgery: the present situation and prospects.

He C, Teng C, Xiong Z, Lin X, Li H, Li X Chin Neurosurg J. 2023; 9(1):14.

PMID: 37170383 PMC: 10176793. DOI: 10.1186/s41016-023-00327-2.


Accuracy of Intracranial Pressure Monitoring-Single Centre Observational Study and Literature Review.

Pelah A, Zakrzewska A, Calviello L, Forcht Dagi T, Czosnyka Z, Czosnyka M Sensors (Basel). 2023; 23(7).

PMID: 37050457 PMC: 10098789. DOI: 10.3390/s23073397.


Invasive Neuromonitoring Modalities in the Pediatric Population.

Lang S, Rahman R, Kumar N, Tucker A, Flanders T, Kirschen M Neurocrit Care. 2023; 38(2):470-485.

PMID: 36890340 PMC: 10269277. DOI: 10.1007/s12028-023-01684-7.


References
1.
Blei A, Olafsson S, Webster S, Levy R . Complications of intracranial pressure monitoring in fulminant hepatic failure. Lancet. 1993; 341(8838):157-8. DOI: 10.1016/0140-6736(93)90016-a. View

2.
Keays R, Alexander G, Williams R . The safety and value of extradural intracranial pressure monitors in fulminant hepatic failure. J Hepatol. 1993; 18(2):205-9. DOI: 10.1016/s0168-8278(05)80247-8. View

3.
JARVIS C, Saxena K . Does prior antibiotic treatment hamper the diagnosis of acute bacterial meningitis? An analysis of a series of 135 childhood cases. Clin Pediatr (Phila). 1972; 11(4):201-4. DOI: 10.1177/000992287201100406. View

4.
Smith R, Alksne J . Infections complicating the use of external ventriculostomy. J Neurosurg. 1976; 44(5):567-70. DOI: 10.3171/jns.1976.44.5.0567. View

5.
Rosner M, Becker D . ICP monitoring: complications and associated factors. Clin Neurosurg. 1976; 23:494-519. DOI: 10.1093/neurosurgery/23.cn_suppl_1.494. View