Accuracy of Intracranial Pressure Monitoring-Single Centre Observational Study and Literature Review

Overview
Journal Sensors (Basel)
Publisher MDPI
Specialty Biotechnology
Date 2023 Apr 13
PMID 37050457
Authors
Affiliations
Soon will be listed here.
Abstract

Intracranial hypertension and adequacy of brain blood flow are primary concerns following traumatic brain injury. Intracranial pressure (ICP) monitoring is a critical diagnostic tool in neurocritical care. However, all ICP sensors, irrespective of design, are subject to systematic and random measurement inaccuracies that can affect patient care if overlooked or disregarded. The wide choice of sensors available to surgeons raises questions about performance and suitability for treatment. This observational study offers a critical review of the clinical and experimental assessment of ICP sensor accuracy and comments on the relationship between actual clinical performance, bench testing, and manufacturer specifications. Critically, on this basis, the study offers guidelines for the selection of ICP monitoring technologies, an important clinical decision. To complement this, a literature review on important ICP monitoring considerations was included. This study utilises illustrative clinical and laboratory material from 1200 TBI patients (collected from 1992 to 2019) to present several important points regarding the accuracy of in vivo implementation of contemporary ICP transducers. In addition, a thorough literature search was performed, with sources dating from 1960 to 2021. Sources considered to be relevant matched the keywords: "intraparenchymal ICP sensors", "fiberoptic ICP sensors", "piezoelectric strain gauge sensors", "external ventricular drains", "CSF reference pressure", "ICP zero drift", and "ICP measurement accuracy". Based on single centre observations and the 76 sources reviewed in this paper, this material reports an overall anticipated measurement accuracy for intraparenchymal transducers of around ± 6.0 mm Hg with an average zero drift of <2.0 mm Hg. Precise ICP monitoring is a key tenet of neurocritical care, and accounting for zero drift is vital. Intraparenchymal piezoelectric strain gauge sensors are commonly implanted to monitor ICP. Laboratory bench testing results can differ from in vivo observations, revealing the shortcomings of current ICP sensors.

Citing Articles

Optical approaches for neurocritical care: Toward non-invasive recording of cerebral physiology in acute brain injury.

Lefebvre A, Steiner N, Rodriguez C, Angelo J, Bar-Kochba E, Mathur R Neurotherapeutics. 2025; 22(1):e00520.

PMID: 39827053 PMC: 11840349. DOI: 10.1016/j.neurot.2024.e00520.


Correlation Between Volume and Pressure of Intracranial Space With Craniectomy Surface Area and Brain Herniation: A Phantom-Based Study.

Sengupta S, Aggarwal R, Singh M Neurotrauma Rep. 2024; 5(1):293-303.

PMID: 38560491 PMC: 10979661. DOI: 10.1089/neur.2024.0006.


Intracranial Pressure Monitoring and Treatment Thresholds in Acute Neural Injury: A Narrative Review of the Historical Achievements, Current State, and Future Perspectives.

Stein K, Froese L, Gomez A, Sainbhi A, Vakitbilir N, Ibrahim Y Neurotrauma Rep. 2023; 4(1):478-494.

PMID: 37636334 PMC: 10457629. DOI: 10.1089/neur.2023.0031.

References
1.
Dhar R, Sandler R, Manwaring K, Kostick N, Mansy H . Noninvasive detection of elevated ICP using spontaneous tympanic membrane pulsation. Sci Rep. 2021; 11(1):21957. PMC: 8578552. DOI: 10.1038/s41598-021-01079-8. View

2.
Khan M, Shallwani H, Khan M, Shamim M . Noninvasive monitoring intracranial pressure - A review of available modalities. Surg Neurol Int. 2017; 8:51. PMC: 5402331. DOI: 10.4103/sni.sni_403_16. View

3.
Ali M, Hashmi M, Shamim S, Salam B, Siraj S, Salim B . Correlation of Optic Nerve Sheath Diameter with Direct Measurement of Intracranial Pressure through an External Ventricular Drain. Cureus. 2019; 11(9):e5777. PMC: 6825501. DOI: 10.7759/cureus.5777. View

4.
Pratt K, Peacock S, Yost K, Freeman W, Collins C, McLaughlin D . Zero-Calibrating External Ventricular Drains: Exploring Practice. J Neurosci Nurs. 2021; 54(1):2-5. DOI: 10.1097/JNN.0000000000000622. View

5.
Czosnyka M, Kirollos R, van Hille P . David Price--Pioneer of digital ICP monitoring, neurosurgeon and teacher. Br J Neurosurg. 2015; 29(3):312-3. DOI: 10.3109/02688697.2015.1012048. View