» Articles » PMID: 10826114

Calibration and LOD/LOQ Estimation of a Chemiluminescent Hybridization Assay for Residual DNA in Recombinant Protein Drugs Expressed in E. Coli Using a Four-parameter Logistic Model

Overview
Publisher Informa Healthcare
Specialty Pharmacology
Date 2000 May 29
PMID 10826114
Citations 2
Authors
Affiliations
Soon will be listed here.
Abstract

Calibration is the process of fitting a model based on reference data points (x, y), then using the model to estimate an unknown x based on a new measured response, y. In DNA assay, x is the concentration, and y is the measured signal volume. A four-parameter logistic model was used frequently for calibration of immunoassay when the response is optical density for enzyme-linked immunosorbent assay (ELISA) or adjusted radioactivity count for radioimmunoassay (RIA). Here, it is shown that the same model or a linearized version of the curve are equally useful for the calibration of a chemiluminescent hybridization assay for residual DNA in recombinant protein drugs and calculation of performance measures of the assay.

Citing Articles

Analytical Validation of a Direct Competitive ELISA for Multiple Mycotoxin Detection in Human Serum.

Garg K, Villavicencio-Aguilar F, Solano-Rivera F, Gilbert L Toxins (Basel). 2022; 14(11).

PMID: 36355977 PMC: 9694295. DOI: 10.3390/toxins14110727.


Nonlinear Regression Improves Accuracy of Characterization of Multiplexed Mass Spectrometric Assays.

Galitzine C, Egertson J, Abbatiello S, Henderson C, Pino L, MacCoss M Mol Cell Proteomics. 2018; 17(5):913-924.

PMID: 29438992 PMC: 5930407. DOI: 10.1074/mcp.RA117.000322.