» Articles » PMID: 10619923

The Effect of Electrolyzed Strong Acid Aqueous Solution on Hemodialysis Equipment

Overview
Journal Artif Organs
Date 2000 Jan 5
PMID 10619923
Citations 10
Authors
Affiliations
Soon will be listed here.
Abstract

We reported the high effectiveness of electrolyzed strong acid aqueous solution (ESAAS) in cleaning hemodialysis lines. Although ESAAS has a strong bactericidal action, one concern is its strong acidity. It has a pH of 2.3-2.7, more than 1,000 mV in oxidation-reduction potential (ORP), and 10-50 ppm of available chlorine. The possibility of metal corrosion, degradation of synthetic resins, chorine gas emission, or dissolving calcium carbonate (CaCO3) deposits due to ESAAS's acidity was tested using in vitro experiments. The bactericidal and antiviral effects of various ESAAS's were also tested. Metal corrosion and synthetic resin degradation, although they occurred, were not serious. There were no problems with chlorine gas emission and dissolving of CaCO3 deposits. Each type of ESAAS showed almost the same bactericidal and antiviral effect, but in some cases differences were observed.

Citing Articles

Antibiofilm activity of electrochemically activated water (ECAW) in the control of Salmonella Heidelberg biofilms on industrial surfaces.

Wilsmann D, Furian T, Carvalho D, Zottis Chitolina G, Lucca V, Emery B Braz J Microbiol. 2023; 54(3):2035-2045.

PMID: 37184738 PMC: 10485189. DOI: 10.1007/s42770-023-01005-2.


Virucidal Activities of Acidic Electrolyzed Water Solutions with Different pH Values against Multiple Strains of SARS-CoV-2.

Takeda Y, Nikaido M, Jamsransuren D, Matsuda S, Ogawa H Appl Environ Microbiol. 2022; 89(1):e0169922.

PMID: 36511659 PMC: 9888296. DOI: 10.1128/aem.01699-22.


Kinetic Modeling of Slightly Acidic Electrolyzed Water Decay Characteristics in Fresh Cabbage Disinfection Against Human Norovirus.

Kang M, Park B, Ha J Front Microbiol. 2021; 12:616297.

PMID: 34295310 PMC: 8290341. DOI: 10.3389/fmicb.2021.616297.


Electrolyzed Water Generated On-Site as a Promising Disinfectant in the Dental Office During the COVID-19 Pandemic.

Farah R, Al-Haj Ali S Front Public Health. 2021; 9:629142.

PMID: 33996714 PMC: 8119747. DOI: 10.3389/fpubh.2021.629142.


Acidic electrolyzed water potently inactivates SARS-CoV-2 depending on the amount of free available chlorine contacting with the virus.

Takeda Y, Uchiumi H, Matsuda S, Ogawa H Biochem Biophys Res Commun. 2020; 530(1):1-3.

PMID: 32828268 PMC: 7359810. DOI: 10.1016/j.bbrc.2020.07.029.