» Articles » PMID: 10458350

Comparison of Mechanical Reliability of Original and Enhanced Mentor Alpha I Penile Prosthesis

Overview
Journal J Urol
Publisher Wolters Kluwer
Specialty Urology
Date 1999 Aug 24
PMID 10458350
Citations 16
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: We report the results of a prospective study of 1,381 Mentor Alpha I penile prostheses implanted to treat impotence, and compare original and enhanced penile prosthesis mechanical reliability.

Materials And Methods: The study consisted of 410 original models manufactured before November 1992 and 971 enhanced models manufactured since December 1992. Implants were further stratified as first time (virgin) or replacements of a previous implant. Mechanical failure-free survival rates for the original prosthetic and enhanced models were compared.

Results: The 5-year survival rate increased from 75.3% for the original to 92.6% for the enhanced model overall (log rank p<0.0001), and from 75.3 to 93.6%, respectively, for the virgin implants only (log rank p<0.0001). The estimated failure rate of approximately 5.6% for the original model was fairly consistent during followup, while the significantly lower failure rate of 1.3% for the enhanced model was not. The failure rate of the enhanced model implants was about 0.8% per year during the first 3.5 years and increased to approximately 3.1% per year thereafter.

Conclusions: Our results strongly support the premise that mechanical reliability is superior with the enhanced compared to the original model.

Citing Articles

Interim Results of Rigicon Penile Prosthesis Implantation in a Single Center: The Largest Series in Türkiye.

Karaman M, Guzel R, Kirkik D, Koca O Urol Res Pract. 2025; 50(4):234-239.

PMID: 39873434 PMC: 11883673. DOI: 10.5152/tud.2025.24122.


Management of Post-RALP SUI and ED - What are and What Should we be Doing?.

Pereira T, Rust J, Bernie H Curr Urol Rep. 2024; 26(1):19.

PMID: 39546126 DOI: 10.1007/s11934-024-01249-x.


Short-term revision rate of Rigicon Testi10 testicular prosthesis in adolescents and adults: a retrospective chart review.

Atwater B, Kirkik D, Wilson S, Koca O, Purohit R, Munoz Vera C Int J Impot Res. 2024; .

PMID: 38714783 DOI: 10.1038/s41443-024-00893-8.


Implant practice building: tenets to growing a successful and rewarding practice.

Wen L, Christine B, Kohler T, Wilson S Int J Impot Res. 2020; 35(5):419-427.

PMID: 33328619 DOI: 10.1038/s41443-020-00384-6.


Prevalence of post-prostatectomy erectile dysfunction and a review of the recommended therapeutic modalities.

Lima T, Bitran J, Frech F, Ramasamy R Int J Impot Res. 2020; 33(4):401-409.

PMID: 33204007 DOI: 10.1038/s41443-020-00374-8.