T2* and Proton Density Measurement of Normal Human Lung Parenchyma Using Submillisecond Echo Time Gradient Echo Magnetic Resonance Imaging

Overview
Journal Eur J Radiol
Specialty Radiology
Date 1999 Jul 10
PMID 10399610
Citations 63
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: To obtain T2* and proton density measurements of normal human lung parenchyma in vivo using submillisecond echo time (TE) gradient echo (GRE) magnetic resonance (MR) imaging.

Materials And Methods: Six normal volunteers were scanned using a 1.5-T system equipped with a prototype enhanced gradient (GE Signa, Waukausha, WI). Images were obtained during breath-holding with acquisition times of 7-16 s. Multiple TEs ranging from 0.7 to 2.5 ms were tested. Linear regression was performed on the logarithmic plots of signal intensity versus TE, yielding measurements of T2* and proton density relative to chest wall muscle. Measurements in supine and prone position were compared, and effects of the level of lung inflation on lung signal were also evaluated.

Results: The signal from the lung parenchyma diminished exponentially with prolongation of TE. The measured T2* in six normal volunteers ranged from 0.89 to 2.18 ms (1.43 +/- 0.41 ms, mean +/- S.D.). The measured relative proton density values ranged between 0.21 and 0.45 (0.29 +/- 0.08, mean +/- S.D.). Calculated T2* values of 1.46 +/- 0.50, 1.01 +/- 0.29 and 1.52 +/- 0.18 ms, and calculated relative proton densities of 0.20 +/- 0.03, 0.32 +/- 0.13 and 0.35 +/- 0.10 were obtained from the anterior, middle and posterior portions of the supine right lung, respectively. The anterior-posterior proton density gradient was reversed in the prone position. There was a pronounced increase in signal from lung parenchyma at maximum expiration compared with maximum inspiration. The ultrashort TE GRE technique yielded images demonstrating signal from lung parenchyma with minimal motion-induced noise.

Conclusion: Quantitative in vivo measurements of lung T2* and relative proton density in conjunction with high-signal parenchymal images can be obtained using a set of very rapid breath-hold images with a recently developed ultrashort TE GRE sequence.

Citing Articles

Recent trends in scientific research in chest radiology: What to do or not to do? That is the critical question in research.

Hatabu H, Yanagawa M, Yamada Y, Hino T, Yamasaki Y, Hata A Jpn J Radiol. 2025; .

PMID: 39815124 DOI: 10.1007/s11604-025-01735-3.


Comparison of weighting algorithms to mitigate respiratory motion in free-breathing neonatal pulmonary radial UTE-MRI.

Gandhi D, Higano N, Hahn A, Gunatilaka C, Torres L, Fain S Biomed Phys Eng Express. 2024; 10(3.

PMID: 38599190 PMC: 11182662. DOI: 10.1088/2057-1976/ad3cdd.


Toward Lung Ventilation Imaging Using Hyperpolarized Diethyl Ether Gas Contrast Agent.

Ariyasingha N, Chowdhury M, Samoilenko A, Salnikov O, Chukanov N, Kovtunova L Chemistry. 2024; 30(25):e202304071.

PMID: 38381807 PMC: 11065616. DOI: 10.1002/chem.202304071.


Echo time-dependent observed T1 and quantitative perfusion in chronic obstructive pulmonary disease using magnetic resonance imaging.

Triphan S, Konietzke M, Biederer J, Eichinger M, Vogelmeier C, Jorres R Front Med (Lausanne). 2024; 10:1254003.

PMID: 38249975 PMC: 10797117. DOI: 10.3389/fmed.2023.1254003.


Shape Anisotropy-Governed High-Performance Nanomagnetosol for In Vivo Magnetic Particle Imaging of Lungs.

Nigam S, Mohapatra J, Makela A, Hayat H, Rodriguez J, Sun A Small. 2023; 20(5):e2305300.

PMID: 37735143 PMC: 10842459. DOI: 10.1002/smll.202305300.