Multichannel Magnetocardiographic Measurements with a Physical Thorax Phantom

Overview
Publisher Springer
Date 1999 Jul 9
PMID 10396834
Citations 4
Authors
Affiliations
Soon will be listed here.
Abstract

Artificial dipolar sources were applied inside a physical thorax phantom to experimentally investigate the accuracy obtainable for non-invasive magnetocardiographic equivalent current dipole localisation. For the measurements, the phantom was filled with saline solution of electrical conductivity 0.21 S m-1. A multichannel cardiomagnetometer was employed to record the magnetic fields generated by seven dipolar sources at distances from 25 mm to 145 mm below the surface of the phantom. The inverse problem was solved using an equivalent current dipole in a homogeneous boundary element torso model. The dipole parameters were determined with a non-linear least squares fitting algorithm. The signal-to-noise ratio (SNR) and the goodness of fit of the calculated localisations were used in assessing the quality of the results. The dependence between the SNR and the goodness of fit was derived, and the results were found to correspond to the model. With SNR between 5 and 10, the average localisation error was found to be 9 +/- 8 mm, while for SNR between 30 and 40 and goodness of fit between 99.5% and 100%, the average error reduced to 3.2 +/- 0.3 mm. The SNR values obtained in this study were also compared with typical clinical values of SNR.

Citing Articles

The magnetocardiogram.

Roth B Biophys Rev (Melville). 2024; 5(2):021305.

PMID: 38827563 PMC: 11139488. DOI: 10.1063/5.0201950.


Characterizing Spatial Signatures of Gastric Electrical Activity Using Biomagnetic Source Localization.

Avci R, Eichler C, Paskaranandavadivel N, Du P, Angeli-Gordon T, Bradshaw L IEEE Trans Biomed Eng. 2022; 69(11):3551-3558.

PMID: 35560086 PMC: 9629986. DOI: 10.1109/TBME.2022.3174847.


Identifying patterns of spatial current dispersion that characterise and separate the Brugada syndrome and complete right-bundle branch block.

Kandori A, Shimizu W, Yokokawa M, Noda T, Kamakura S, Miyatake K Med Biol Eng Comput. 2004; 42(2):236-44.

PMID: 15125155 DOI: 10.1007/BF02344637.


A method for detecting myocardial abnormality by using a total current-vector calculated from ST-segment deviation of a magnetocardiogram signal.

Kandori A, Kanzaki H, Miyatake K, Hashimoto S, Itoh S, Tanaka N Med Biol Eng Comput. 2001; 39(1):21-8.

PMID: 11214269 DOI: 10.1007/BF02345262.

References
1.
Oeff M, Burghoff M . Magnetocardiographic localization of the origin of ventricular ectopic beats. Pacing Clin Electrophysiol. 1994; 17(3 Pt 2):517-22. DOI: 10.1111/j.1540-8159.1994.tb01420.x. View

2.
Hren R, Zhang X, Stroink G . Comparison between electrocardiographic and magnetocardiographic inverse solutions using the boundary element method. Med Biol Eng Comput. 1996; 34(2):110-4. DOI: 10.1007/BF02520014. View

3.
Moshage W, Achenbach S, Gohl K, Bachmann K . Evaluation of the non-invasive localization accuracy of cardiac arrhythmias attainable by multichannel magnetocardiography (MCG). Int J Card Imaging. 1996; 12(1):47-59. DOI: 10.1007/BF01798116. View

4.
Nenonen J, Makijarvi M, Toivonen L, Forsman K, Leinio M, Montonen J . Non-invasive magnetocardiographic localization of ventricular pre-excitation in the Wolff-Parkinson-White syndrome using a realistic torso model. Eur Heart J. 1993; 14(2):168-74. DOI: 10.1093/eurheartj/14.2.168. View

5.
Weismuller P, Abraham-Fuchs K, Schneider S, Richter P, Kochs M, Hombach V . Magnetocardiographic non-invasive localization of accessory pathways in the Wolff-Parkinson-White syndrome by a multichannel system. Eur Heart J. 1992; 13(5):616-22. DOI: 10.1093/oxfordjournals.eurheartj.a060224. View