» Articles » PMID: 9806519

Preliminary Experience with Frameless Stereotactic Radiotherapy

Overview
Specialties Oncology
Radiology
Date 1998 Nov 7
PMID 9806519
Citations 19
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To report initial clinical experience with a novel high-precision stereotactic radiotherapy system.

Methods And Materials: Sixty patients ranging in age from 2 to 82 years received a total of 1426 treatments with the University of Florida frameless stereotactic radiotherapy system. Of the total, 39 (65%) were treated with stereotactic radiotherapy (SRT) alone, and 21 (35%) received SRT as a component of radiotherapy. Pathologic diagnoses included meningiomas (15 patients), low-grade astrocytomas (11 patients), germinomas (9 patients), and craniopharyngiomas (5 patients). The technique was used as means of dose escalation in 11 patients (18%) with aggressive tumors. Treatment reproducibility was measured by comparing bite plate positioning registered by infrared light-emitting diodes (IRLEDs) with the stereotactic radiosurgery reference system, and with measurements from each treatment arc for the 1426 daily treatments (5808 positions). We chose 0.3 mm vector translation error and 0.3 degrees rotation about each axis as the maximum tolerated misalignment before treating each arc.

Results: With a mean follow-up of 11 months, 3 patients had recurrence of malignant disease. Acute side effects were minimal. Of 11 patients with low grade astrocytomas, 4 (36%) had cerebral edema and increased enhancement on MR scans in the first year, and 2 required steroids. All had resolution and marked tumor involution on follow-up imaging. Bite plate reproducibility was as follows. Translational errors: anterior-posterior, 0.01 +/- 0.10; lateral, 0.02 +/- 0.07; axial, 0.01 +/- 0.10. Rotational errors (degrees): anterior-posterior, 0.00 +/- 0.03; lateral, 0.00 +/- 0.06; axial, 0.01 +/- 0.04. No patient treatment was delivered beyond the maximum tolerated misalignment. Daily treatment was delivered in approximately 15 min per patient.

Conclusion: Our initial experience with stereotactic radiotherapy using the infrared camera guidance system was good. Patient selection and treatment strategies are evolving rapidly. Treatment accuracy was the best reported, and the treatment approach was practical.

Citing Articles

Advances in Image-Guided Radiotherapy in the Treatment of Oral Cavity Cancer.

Nien H, Wang L, Liao L, Lin P, Wu C, Shueng P Cancers (Basel). 2022; 14(19).

PMID: 36230553 PMC: 9561985. DOI: 10.3390/cancers14194630.


Development of a Real-Time Thermoplastic Mask Compression Force Monitoring System Using Capacitive Force Sensor.

Kim T, Cho M, Shin D, Shin D, Kim S Front Robot AI. 2022; 9:778594.

PMID: 35875702 PMC: 9298856. DOI: 10.3389/frobt.2022.778594.


Analysis of intrafraction motion in CyberKnife-based stereotaxy using mask based immobilization and 6D-skull tracking.

Kataria T, Narang K, Gupta D, Bisht S, Abhishek A, Goyal S J Radiosurg SBRT. 2018; 4(3):203-212.

PMID: 29296445 PMC: 5658803.


Radiosurgery technology development and use.

Meeks S, Pukala J, Ramakrishna N, Willoughby T, Bova F J Radiosurg SBRT. 2018; 1(1):21-29.

PMID: 29296294 PMC: 5658896.


Treatment of MRI-Diagnosed Trigeminal Peripheral Nerve Sheath Tumors by Stereotactic Radiotherapy in Dogs.

Hansen K, Zwingenberger A, Theon A, Pfeiffer I, Kent M J Vet Intern Med. 2016; 30(4):1112-20.

PMID: 27279132 PMC: 5089643. DOI: 10.1111/jvim.13970.