Multicenter Comparison of BACTEC 9050 and BACTEC 9240 Blood Culture Systems

Overview
Specialty Microbiology
Date 1998 Jun 10
PMID 9620384
Citations 3
Authors
Affiliations
Soon will be listed here.
Abstract

The overall recovery of organisms and time to detection with the BACTEC 9050 and BACTEC 9240 systems were compared in a multicenter evaluation. In the first phase of the study, a total of 4,383 compliant aerobic (Plus Aerobic/F) blood culture sets were processed. There was no significant difference in the recovery of individual groups of organisms with the two systems, with the exception of Streptococcus pneumoniae which was isolated more frequently with BACTEC 9050. False-positive signals occurred more often with BACTEC 9240 (58 cultures) than with BACTEC 9050 (43 cultures), but false-negative cultures were uncommon with both systems (3 cultures for each system). Time to detection of positive cultures of clinically significant organisms was essentially the same with both instruments. In the second phase of the study, 2,431 compliant anaerobic (Plus Anaerobic/F) blood culture sets were processed. There was no significant difference in the recovery of organisms with BACTEC 9050 compared with BACTEC 9240. Significantly (P < 0.03) more false-positive signals occurred with BACTEC 9240 (15 cultures) than with BACTEC 9050 (4 cultures). Likewise, more false-negative cultures occurred with BACTEC 9240 (11 cultures) than with BACTEC 9050 (8 cultures). Time to detection of positive cultures of clinically significant organisms was essentially the same with both systems with the exception of anaerobes (N = 10), which were recovered earlier (P < 0.01) with BACTEC 9240 (35.0 h) than with BACTEC 9050 (61.4 h).

Citing Articles

MALDI-TOF MS performance compared to direct examination, culture, and 16S rDNA PCR for the rapid diagnosis of bone and joint infections.

Lallemand E, Coiffier G, Arvieux C, Brillet E, Guggenbuhl P, Jolivet-Gougeon A Eur J Clin Microbiol Infect Dis. 2016; 35(5):857-66.

PMID: 26942744 DOI: 10.1007/s10096-016-2608-x.


Comparison of the VersaTREK blood culture system against the Bactec9240 system in patients with suspected bloodstream infections.

Dreyer A, Ismail N, Nkosi D, Lindeque K, Matthews M, Van Zyl D Ann Clin Microbiol Antimicrob. 2011; 10:4.

PMID: 21294908 PMC: 3042901. DOI: 10.1186/1476-0711-10-4.


Time to positivity in blood cultures of adults with Streptococcus pneumoniae bacteremia.

Peralta G, Rodriguez-Lera M, Garrido J, Ansorena L, Roiz M BMC Infect Dis. 2006; 6:79.

PMID: 16643662 PMC: 1475865. DOI: 10.1186/1471-2334-6-79.

References
1.
Nolte F, Williams J, Jerris R, MORELLO J, Leitch C, Matushek S . Multicenter clinical evaluation of a continuous monitoring blood culture system using fluorescent-sensor technology (BACTEC 9240). J Clin Microbiol. 1993; 31(3):552-7. PMC: 262818. DOI: 10.1128/jcm.31.3.552-557.1993. View

2.
Smith J, Bryce E, Roberts F . Comparison of BACTEC 9240 and BacT/Alert blood culture systems in an adult hospital. J Clin Microbiol. 1995; 33(7):1905-8. PMC: 228295. DOI: 10.1128/jcm.33.7.1905-1908.1995. View

3.
Cockerill 3rd F, Reed G, Hughes J, Torgerson C, Vetter E, Harmsen W . Clinical comparison of BACTEC 9240 plus aerobic/F resin bottles and the isolator aerobic culture system for detection of bloodstream infections. J Clin Microbiol. 1997; 35(6):1469-72. PMC: 229769. DOI: 10.1128/jcm.35.6.1469-1472.1997. View

4.
Pohlman J, Kirkley B, Easley K, Basille B, Washington J . Controlled clinical evaluation of BACTEC Plus Aerobic/F and BacT/Alert Aerobic FAN bottles for detection of bloodstream infections. J Clin Microbiol. 1995; 33(11):2856-8. PMC: 228594. DOI: 10.1128/jcm.33.11.2856-2858.1995. View

5.
Jorgensen J, Mirrett S, McDonald L, Murray P, Weinstein M, Fune J . Controlled clinical laboratory comparison of BACTEC plus aerobic/F resin medium with BacT/Alert aerobic FAN medium for detection of bacteremia and fungemia. J Clin Microbiol. 1997; 35(1):53-8. PMC: 229511. DOI: 10.1128/jcm.35.1.53-58.1997. View