The Effect of Activity Outside the Direct Field of View in a 3D-only Whole-body Positron Tomograph

Overview
Journal Phys Med Biol
Publisher IOP Publishing
Date 1998 May 8
PMID 9572513
Citations 5
Authors
Affiliations
Soon will be listed here.
Abstract

The ECAT EXACT3D (CTI/Siemens 966) 3D-only PET tomograph has unprecedented sensitivity due to the large BGO (bismuth germanate) detector volume. However, the consequences of a large (23.4 cm) axial field-of-view (FOV) and the need for a patient port diameter to accommodate body scanning make the device more sensitive to photons arising from activity outside the direct (coincidence) FOV. This leads to relatively higher deadtime and an increased registration of random and scatter (true) coincidences. The purpose of this study is to determine the influence of activity outside the FOV on (i) noise-equivalent counts (NEC) and (ii) the performance of a 'model-based' scatter correction algorithm, and to investigate the effect of side shielding additional to that supplied with the tomograph. Annular shielding designed for brain scanning increased the NEC for blood flow (H[2]15O) measurement (integrated over 120 s) by up to 25%. For 11C tracer studies, the increase is less than 5% over 120 min. Purpose-built additional body shielding, made to conform to the shape of a volunteer, reduced the randoms count rate in a heart blood flow measurement (H[2]15O) by about 30%. After scatter correction the discrepancy between ROI count ratios for compartments within the 20 cm diameter 'Utah' phantom differed by less than 5% from true (sampled) activity concentration ratios. This was so with or without activity outside the FOV and with or without additional side shielding. Count rate performance is thus improved by extra shielding but more improvement is seen in head than in body scanning. Measurement of heart blood flow using bolus injections of H(2)15O would benefit from the use of detectors with lower deadtime and superior timing resolution such as LSO (lutetium oxyorthosilicate).

Citing Articles

Quantification of brain oxygen extraction and metabolism with [O]-gas PET: A technical review in the era of PET/MRI.

Fan A, An H, Moradi F, Rosenberg J, Ishii Y, Nariai T Neuroimage. 2020; 220:117136.

PMID: 32634594 PMC: 7592419. DOI: 10.1016/j.neuroimage.2020.117136.


Whole-body biodistribution and the influence of body activity on brain kinetic analysis of the C-PiB PET scan.

Akamatsu G, Nishio T, Adachi K, Ikari Y, Senda M Radiol Phys Technol. 2017; 10(4):464-474.

PMID: 28895034 DOI: 10.1007/s12194-017-0419-0.


Effects of a novel tungsten-impregnated rubber neck shield on the quality of cerebral images acquired using O-labeled gas.

Wagatsuma K, Oda K, Miwa K, Inaji M, Sakata M, Toyohara J Radiol Phys Technol. 2017; 10(4):422-430.

PMID: 28823084 DOI: 10.1007/s12194-017-0414-5.


Quantification of ligand PET studies using a reference region with a displaceable fraction: application to occupancy studies with [(11)C]-DASB as an example.

Turkheimer F, Selvaraj S, Hinz R, Murthy V, Bhagwagar Z, Grasby P J Cereb Blood Flow Metab. 2011; 32(1):70-80.

PMID: 21811290 PMC: 3323353. DOI: 10.1038/jcbfm.2011.108.


Study of an image-derived SUV and a modified SUV using mouse FDG-PET.

Zheng X, Yu C, Sha W, Radu C, Huang S, Feng D Nucl Med Biol. 2011; 38(3):353-62.

PMID: 21492784 PMC: 3078337. DOI: 10.1016/j.nucmedbio.2010.10.002.