Cartilage and Subchondral Bone Thickness Distribution with MR Imaging

Overview
Journal Acad Radiol
Specialty Radiology
Date 1998 Jan 27
PMID 9442203
Citations 4
Authors
Affiliations
Soon will be listed here.
Abstract

Rationale And Objectives: The authors mapped articular cartilage (AC) and subchondral bone (SB) thicknesses in human acetabula in vitro by using magnetic resonance (MR) imaging and validated AC measurements by using light microscopy.

Materials And Methods: Left and right acetabula from a deceased patient who had undergone left hemiarthroplasty were imaged with fat-suppressed spoiled gradient-recalled acquisition in the steady state (repetition time = 55 msec, echo time = 15 msec, flip angle = 50 degrees, matrix = 256 x 256, field of view = 8 cm). AC and SB thickness maps were generated from image data by using analytic geometry, which enabled correction for thickness overestimation due to oblique sectioning. Cartilage bone plugs were extracted from the acetabula, and light microscopy was used to validate the thickness measurements obtained with MR imaging.

Results: Standard errors between thickness measurements obtained with MR imaging and light microscopy were 0.37 and 0.33 mm for the left and right AC, respectively, which is consistent with the voxel resolution of the MR imaging sequence (0.31 x 0.31 x 0.8 mm). SB thickness of the cartilage plugs could not be reliably measured with light microscopy and, therefore, could not be validated. Contour maps showed that SB thickness gradients were rapid and focal compared with the rather smooth gradients in AC thickness; however, thicker AC was accompanied by thicker SB for left (r2 = .261, P = .0001) and right (r2 = .308, P = .0001) acetabula. Average thickness differences between left and right acetabular AC and SB were 0.13 mm (P = .015) and 0.11 mm (P = .026), respectively. Although it was the operated hip that had thicker articular tissues, the differences were within the pixel resolution (< 0.31 mm).

Conclusion: AC and SB thickness distribution can be accurately determined by combining noninvasive MR imaging and analytic geometry, which may also provide a means for quantitative, longitudinal assessment of focal AC defects.

Citing Articles

Measurement of the "safe zone" and the "dangerous zone" for the screw placement on the quadrilateral surface in the treatment of pelvic and acetabular fractures with Stoppa approach by computational 3D technology.

Zhang S, Su W, Luo Q, Leung F, Chen B Biomed Res Int. 2014; 2014:386950.

PMID: 24605328 PMC: 3925522. DOI: 10.1155/2014/386950.


MRI of the temporo-mandibular joint: which sequence is best suited to assess the cortical bone of the mandibular condyle? A cadaveric study using micro-CT as the standard of reference.

Karlo C, Patcas R, Kau T, Watzal H, Signorelli L, Muller L Eur Radiol. 2012; 22(7):1579-85.

PMID: 22322312 DOI: 10.1007/s00330-012-2391-8.


The effects of exercise on human articular cartilage.

Eckstein F, Hudelmaier M, Putz R J Anat. 2006; 208(4):491-512.

PMID: 16637874 PMC: 2100201. DOI: 10.1111/j.1469-7580.2006.00546.x.


Subchondral bone and cartilage thickness from MRI: effects of chemical-shift artifact.

McGibbon C, Bencardino J, Palmer W MAGMA. 2003; 16(1):1-9.

PMID: 12695880 DOI: 10.1007/s10334-003-0001-0.