A Comparative Laboratory Study of Liquid Nitrogen and Argon Gas Cryosurgery Systems

Overview
Journal Cryobiology
Publisher Elsevier
Specialty Biology
Date 1998 Jan 13
PMID 9425653
Citations 11
Authors
Affiliations
Soon will be listed here.
Abstract

Cryotherapy can now be applied using a variety of delivery systems and cryogens. We compared the Cryotech LCS 3000 liquid nitrogen system (Spembly, Andover, UK) with the CRYOcare argon gas-based system (Irvine, CA, U.S.A.) using three different 3-mm cryoprobes: an old liquid nitrogen probe (N-probe), a new N-probe featuring gas bypass and an argon gas probe. Each probe was tested in two models: (i) fresh sheep liver at 20 degrees C--the probe was inserted to a depth of 1.5 cm; the rate of ice ball formation was monitored by recording radial temperatures every 15 s at 5, 10, 15, and 20 mm from the cryoprobe, and the ice-ball diameter was measured every 2.5 min. After 10 min, the probe was warmed and the time taken until it could be extracted from the liver was recorded. (ii) Warm water bath--the probe was immersed in warm water (42 degrees C) for 15 min and the ice-ball diameter was measured at 5-min intervals. Radial temperatures in liver declined more rapidly (P < 0.001) and time to probe extraction was less (P < 0.01) when the argon gas system was used. The new N-probe performed better than its older counterpart, but was still slower than the argon gas system. In liver (20 degrees C), ice-ball diameters were similar after 10 min, but in warm water, they were larger when the new N-probe was used (P < 0.02). It would appear that the argon gas system is initially faster, but it does not achieve as large an ice ball in a warm environment as the liquid nitrogen system.

Citing Articles

Co-ablation versus cryoablation for the treatment of stage III-IV non-small cell lung cancer: A prospective, noninferiority, randomized, controlled trial (RCT).

Yang W, An Y, Li Q, Liu C, Zhu B, Huang Q Thorac Cancer. 2020; 12(4):475-483.

PMID: 33319493 PMC: 7882381. DOI: 10.1111/1759-7714.13779.


Liquid Nitrogen-Based Cryoablation in In Vivo Porcine Tissue: A Pilot Study.

Chang D, Mohan P, Amin A, Garcia-Buitrago M, Rodriguez J, Peaden R Asian Pac J Cancer Prev. 2020; 21(10):3069-3075.

PMID: 33112569 PMC: 7798152. DOI: 10.31557/APJCP.2020.21.10.3069.


Gel phantom study of a cryosurgical probe with a thermosiphon effect and liquid nitrogen-cooled aluminum thermal storage blocks.

Isoda H, Takehara Y, Fujino H, Sone K, Suzuki T, Tsuzaki Y Nagoya J Med Sci. 2015; 77(3):399-407.

PMID: 26412886 PMC: 4574327.


Renal ablation update.

Khiatani V, Dixon R Semin Intervent Radiol. 2014; 31(2):157-66.

PMID: 25049445 PMC: 4078154. DOI: 10.1055/s-0034-1373790.


Lower recurrence rate in chondroblastoma using extended curettage and cryosurgery.

Mashhour M, Abdel Rahman M Int Orthop. 2013; 38(5):1019-24.

PMID: 24248272 PMC: 3997782. DOI: 10.1007/s00264-013-2178-9.