» Articles » PMID: 9420669

A Method of Simultaneous Dual-radionuclide Cardiac Imaging with Technetium 99m and Thallium 201. I: Analysis of Interradionuclide Crossover and Validation in Phantoms

Overview
Journal J Nucl Cardiol
Date 1994 Jan 1
PMID 9420669
Citations 5
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Simultaneous dual-radionuclide technetium 99m/thallium 201 scintigraphy can potentially produce perfectly aligned stress and rest images in less time than conventional protocols. However, interradionuclide crossover limits diagnostic accuracy. Accordingly, we evaluated 99mTc and 201Tl crossover in line and heart phantoms.

Methods And Results: 99mTc crossover in the 201Tl imaging window constituted as much as one half of the counts in the 99mTc window, varied significantly with attenuation, and was spatially incoherent. 201Tl crossover was relatively small, less variable, and spatially similar to the primary image. Based on these findings, the following simultaneous dual-radionuclide 99mTc/201Tl method was developed, and validated in line and heart phantoms. The 99mTc source is imaged first into dual 201Tl/99mTc windows, followed by 201Tl administration and dual-radionuclide imaging. The single-radionuclide 99mTc image in the 201Tl window is count-normalized for acquisition time and then subtracted from the dual-radionuclide 201Tl image to specifically correct for 99mTc crossover. Image quality of the corrected dual-radionuclide 201Tl images approached their single-radionuclide counterparts. Correction for 201Tl crossover was relatively unimportant.

Conclusion: Simultaneous dual-radionuclide 99mTc/201Tl myocardial scintigraphy is feasible with 99mTc crossover correction specific to each acquisition. The proposed dual-radionuclide 99mTc/201Tl method and the principles on which it is based can be applied to a broad range of dual-radionuclide pairs.

Citing Articles

Iterative deconvolution of simultaneous 99mTc and 201Tl projection data measured on a CdZnTe-based cardiac SPECT scanner.

Kacperski K, Erlandsson K, Ben-Haim S, Hutton B Phys Med Biol. 2011; 56(5):1397-414.

PMID: 21317483 PMC: 3111906. DOI: 10.1088/0031-9155/56/5/012.


Preliminary clinical results of photon energy recovery in simultaneous rest Tl-201/stress Tc-99m sestamibi myocardial SPECT.

Hannequin P, Weinmann P, Mas J, Vinot S J Nucl Cardiol. 2001; 8(2):144-51.

PMID: 11295691 DOI: 10.1067/mnc.2001.111799.


Clinical value of triple-energy window scatter correction in simultaneous dual-isotope single-photon emission tomography with 123I-BMIPP and 201Tl.

Yang J, Yamamoto K, Sadato N, Tsuchida T, Takahashi N, Hayashi N Eur J Nucl Med. 1997; 24(9):1099-106.

PMID: 9283101 DOI: 10.1007/BF01254240.


Quantitative comparison of single-isotope and dual-isotope stress-rest single-photon emission computed tomographic imaging for reversibility of defects.

Siebelink H, Natale D, Sinusas A, Wackers F J Nucl Cardiol. 1996; 3(6 Pt 1):483-93.

PMID: 8989673 DOI: 10.1016/s1071-3581(96)90058-6.


Simultaneous dual-radionuclide imaging: are the images trustworthy?.

Links J Eur J Nucl Med. 1996; 23(10):1289-91.

PMID: 8781131 DOI: 10.1007/BF01367582.

References
1.
Jaszczak R, Greer K, Floyd Jr C, Harris C, Coleman R . Improved SPECT quantification using compensation for scattered photons. J Nucl Med. 1984; 25(8):893-900. View

2.
Kessler C, Reuther R, Kimmig B, Pietzsch T . Dual isotope scintigraphy in stroke patients. Neuroradiology. 1984; 26(2):113-7. DOI: 10.1007/BF00339858. View

3.
Devous Sr M, Payne J, Lowe J . Dual-isotope brain SPECT imaging with technetium-99m and iodine-123: clinical validation using xenon-133 SPECT. J Nucl Med. 1992; 33(11):1919-24. View

4.
Zimmermann R, Rauch B, Kapp M, Bubeck B, Neumann F, Seitz F . Myocardial scintigraphy with iodine-123 phenylpentadecanoic acid and thallium-201 in patients with coronary artery disease: a comparative dual-isotope study. Eur J Nucl Med. 1992; 19(11):946-54. DOI: 10.1007/BF00175860. View

5.
Cohn J, Burke L . Nitroprusside. Ann Intern Med. 1979; 91(5):752-7. DOI: 10.7326/0003-4819-91-5-752. View