Technical Accuracy of a Neuronavigation System Measured with a High-precision Mechanical Micromanipulator

Overview
Journal Neurosurgery
Specialty Neurosurgery
Date 1997 Dec 24
PMID 9402599
Citations 9
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: This study was designed to determine and evaluate the different system-inherent sources of erroneous target localization of a light-emitting diode (LED)-based neuronavigation system (StealthStation, Stealth Technologies, Boulder, CO).

Methods: The localization accuracy was estimated by applying a high-precision mechanical micromanipulator to move and exactly locate (+/- 0.1 micron) the pointer at multiple positions in the physical three-dimensional space. The localization error was evaluated by calculating the spatial distance between the (known) LED positions and the LED coordinates measured by the neuronavigator. The results are based on a study of approximately 280,000 independent coordinate measurements.

Results: The maximum localization error detected was 0.55 +/- 0.29 mm, with the z direction (distance to the camera array) being the most erroneous coordinate. Minimum localization error was found at a distance of 1400 mm from the central camera (optimal measurement position). Additional error due to 1) mechanical vibrations of the camera tripod (+/- 0.15 mm) and the reference frame (+/- 0.08 mm) and 2) extrapolation of the pointer tip position from the LED coordinates of at least +/- 0.12 mm were detected, leading to a total technical error of 0.55 +/- 0.64 mm.

Conclusions: Based on this technical accuracy analysis, a set of handling recommendations is proposed, leading to an improved localization accuracy. The localization error could be reduced by 0.3 +/- 0.15 mm by correct camera positioning (1400 mm distance) plus 0.15 mm by vibration-eliminating fixation of the camera. Correct handling of the probe during the operation may improve the accuracy by up to 0.1 mm.

Citing Articles

A method for efficient, rapid, and minimally invasive implantation of individual non-functional motes with penetrating subcellular-diameter carbon fiber electrodes into rat cortex.

Letner J, Lam J, Copenhaver M, Barrow M, Patel P, Richie J bioRxiv. 2025; .

PMID: 39974888 PMC: 11838573. DOI: 10.1101/2025.02.05.636655.


Augmented Reality in Neurosurgery.

van Doormaal J, van Doormaal T Adv Exp Med Biol. 2024; 1462:351-374.

PMID: 39523276 DOI: 10.1007/978-3-031-64892-2_21.


Navigation Guided Biopsy Is as Effective as Frame-Based Stereotactic Biopsy.

Lim D, Kim S, Na Y, Cho J J Pers Med. 2023; 13(5).

PMID: 37240878 PMC: 10219353. DOI: 10.3390/jpm13050708.


Mapping and Preserving the Visuospatial Network by repetitive nTMS and DTI Tractography in Patients With Right Parietal Lobe Tumors.

Raffa G, Quattropani M, Marzano G, Curcio A, Rizzo V, Sebestyen G Front Oncol. 2021; 11:677172.

PMID: 34249716 PMC: 8268025. DOI: 10.3389/fonc.2021.677172.


Rapid fabrication of custom patient biopsy guides.

Rajon D, Bova F, Chi Y, Friedman W J Appl Clin Med Phys. 2009; 10(4):260-272.

PMID: 19918219 PMC: 5720585. DOI: 10.1120/jacmp.v10i4.2897.