Real-time Assessment of the Parameter of Nonlinearity in Tissue Using "nonlinear Shadowing"

Overview
Specialty Radiology
Date 1996 Jan 1
PMID 9123646
Citations 2
Authors
Affiliations
Soon will be listed here.
Abstract

A real-time pulse-echo method for assessing tissue nonlinearity is presented. Here, tissue nonlinearity refers to the parameter of nonlinearity, B/A, and not the frequency dependence of the attenuation. Our nonlinearity assessment method is based on two pulse transmissions per beam position. One pulse is transmitted at a high intensity that propagates nonlinearily and, the other, at a low level, that propagates linearly. Echo data are used to estimate a function representing the decay of fundamental frequency amplitude due to nonlinearity along the ultrasound path. This function is then mapped into an image that includes shadows which relate to the nonlinearity of the preceding regions. To delineate a region of interest from the background tissue, a normalization curve is defined and used to compensate for the nonlinearity of the background tissue. Experiments are performed on a tissue-mimicking phantom and in vitro tissue samples including regions of high nonlinearity. Results indicate that the method can effectively identify nonlinear regions and assess the extent of the nonlinearity.

Citing Articles

Effects of acoustic nonlinearity on pulse-echo attenuation coefficient estimation from tissue-mimicking phantoms.

Coila A, Oelze M J Acoust Soc Am. 2020; 148(2):805.

PMID: 32873024 PMC: 7434466. DOI: 10.1121/10.0001690.


Effects of acoustic nonlinearities on the ultrasonic backscatter coefficient estimation.

Coila A, Oelze M J Acoust Soc Am. 2019; 146(1):85.

PMID: 31370607 PMC: 6615998. DOI: 10.1121/1.5115355.