Empirical and Theoretical Investigation of the Noise Performance of Indirect Detection, Active Matrix Flat-panel Imagers (AMFPIs) for Diagnostic Radiology

Overview
Journal Med Phys
Specialty Biophysics
Date 1997 Jan 1
PMID 9029542
Citations 63
Authors
Affiliations
Soon will be listed here.
Abstract

Noise properties of active matrix, flat-panel imagers under conditions relevant to diagnostic radiology are investigated. These studies focus on imagers based upon arrays with pixels incorporating a discrete photodiode coupled to a thin-film transistor, both fabricated from hydrogenated amorphous silicon. These optically sensitive arrays are operated with an overlying x-ray converter to allow indirect detection of incident x rays. External electronics, including gate driver circuits and preamplification circuits, are also required to operate the arrays. A theoretical model describing the signal and noise transfer properties of the imagers under conditions relevant to diagnostic radiography, fluoroscopy, and mammography is developed. This frequency-dependent model is based upon a cascaded systems analysis wherein the imager is conceptually divided into a series of stages having intrinsic gain and spreading properties. Predictions from the model are compared with x-ray sensitivity and noise measurements obtained from individual pixels from an imager with a pixel format of 1536 x 1920 pixels at a pixel pitch of 127 microns. The model is shown to be in excellent agreement with measurements obtained with diagnostic x rays using various phosphor screens. The model is used to explore the potential performance of existing and hypothetical imagers for application in radiography, fluoroscopy, and mammography as a function of exposure, additive noise, and fill factor. These theoretical predictions suggest that imagers of this general design incorporating a CsI: Tl intensifying screen can be optimized to provide detective quantum efficiency (DQE) superior to existing screen-film and storage phosphor systems for general radiography and mammography. For fluoroscopy, the model predicts that with further optimization of a-Si:H imagers, DQE performance approaching that of the best x-ray image intensifier systems may be possible. The results of this analysis suggest strategies for future improvements of this imaging technology.

Citing Articles

Examining the influence of digital phantom models in virtual imaging trials for tomographic breast imaging.

Kavuri A, Das M J Med Imaging (Bellingham). 2025; 12(1):015501.

PMID: 39744152 PMC: 11686409. DOI: 10.1117/1.JMI.12.1.015501.


Radiation Detectors and Sensors in Medical Imaging.

Michail C, Liaparinos P, Kalyvas N, Kandarakis I, Fountos G, Valais I Sensors (Basel). 2024; 24(19).

PMID: 39409289 PMC: 11478476. DOI: 10.3390/s24196251.


Maximizing microcalcification detectability in low-dose dedicated cone-beam breast CT: parallel cascades-based theoretical analysis.

Larsen T, Tseng H, Trinate R, Fu Z, Chiang J, Karellas A J Med Imaging (Bellingham). 2024; 11(3):033501.

PMID: 38756437 PMC: 11095120. DOI: 10.1117/1.JMI.11.3.033501.


Noise power spectrum (NPS) in computed tomography: Enabling local NPS measurement without stationarity and ergodicity assumptions.

Zhang C, Li K, Zhang R, Chen G Med Phys. 2024; 51(7):4655-4672.

PMID: 38709982 PMC: 11233243. DOI: 10.1002/mp.17112.


Experimental measurement of local noise power spectrum (NPS) in photon counting detector-CT (PCD-CT) using a single data acquisition.

Zhang C, Li K, Zhang R, Chen G Med Phys. 2024; 51(6):4081-4094.

PMID: 38703355 PMC: 11147724. DOI: 10.1002/mp.17110.