» Articles » PMID: 8912782

Improved Arterial Oxygenation After Oleic Acid Lung Injury in the Pig Using a Computer-controlled Mechanical Ventilator

Overview
Specialty Critical Care
Date 1996 Nov 1
PMID 8912782
Citations 29
Authors
Affiliations
Soon will be listed here.
Abstract

We compared computer-controlled mechanical ventilation programmed for biologic variability of respiratory rate (RR) and tidal volume (VT) with conventional intermittent positive-pressure ventilation (IPPV) in an oleic acid (OA) lung injury model. Seventeen pigs were ventilated with an Ohio 7000 anesthesia ventilator. Minute ventilation (VE) was adjusted to maintain PaCO2 at 30 to 35 mm Hg at baseline and was not altered further. OA was infused at 0.2 ml/kg/h until PaO2 decreased to < 125 mm Hg (F(I)O2 = 0.5). Animals were randomly assigned to continue with conventional IPPV (control group; n = 8) or had IPPV computer-controlled (computer group; n = 9). Hemodynamic, respiratory gas, airway pressure, and volume data were obtained at baseline (before OA infusion), at Time 30 (after infusion), and at 30-min intervals for 240 min after OA. At experiment completion, the lungs were removed to determine the wet:dry weight ratios. The control group had RR fixed at 20 breaths/min. The computer group had a RR of 20 +/- 2.3 breaths/min (range, 15 to 27 breaths/min), comprising 369 different RR values with reciprocal changes in VT over 1,089 s before the program looped to repeat itself. There was no difference between groups in the volume of OA infused. By 120 min after lung injury, animals in the computer group had significantly greater PaO2, associated with a lower Qs/QT. Mean airway pressures and mean peak airway pressures were not different in the two groups. By 180 min, respiratory system compliance (Crs) was significantly lower in the control group. The wet:dry lung weight ratios were greater in the control group. Thus, in a porcine model of OA lung injury, computer-controlled mechanical ventilation, which is programmed for biologic variability, resulted in improved blood oxygenation without increasing mean airway pressures when compared with conventional IPPV.

Citing Articles

Oscillatory ventilation enhances oxygenation and reduces inflammation in an animal model of acute respiratory distress syndrome: an experimental study.

Forgiarini Junior L, Forgiarini L, Paludo A, Mariano R, de Moraes M, Felix E Braz J Anesthesiol. 2024; 75(1):844576.

PMID: 39586375 PMC: 11731209. DOI: 10.1016/j.bjane.2024.844576.


Variable positive end-expiratory pressure in an experimental model of acute respiratory distress syndrome: an advanced ventilation modality.

Sudy R, Diaper J, Bizzotto D, Dellaca R, Petak F, Habre W BJA Open. 2024; 11:100302.

PMID: 39224447 PMC: 11367518. DOI: 10.1016/j.bjao.2024.100302.


Pink noise promotes sooner state transitions during bimanual coordination.

Brink K, Kim S, Sommerfeld J, Amazeen P, Stergiou N, Likens A Proc Natl Acad Sci U S A. 2024; 121(31):e2400687121.

PMID: 39042677 PMC: 11294992. DOI: 10.1073/pnas.2400687121.


The Constrained Disorder Principle Accounts for the Variability That Characterizes Breathing: A Method for Treating Chronic Respiratory Diseases and Improving Mechanical Ventilation.

Adar O, Hollander A, Ilan Y Adv Respir Med. 2023; 91(5):350-367.

PMID: 37736974 PMC: 10514877. DOI: 10.3390/arm91050028.


Phrenic nerve stimulation mitigates hippocampal and brainstem inflammation in an ARDS model.

Bassi T, Rohrs E, Fernandez K, Ornowska M, Nicholas M, Wittmann J Front Physiol. 2023; 14:1182505.

PMID: 37215178 PMC: 10196250. DOI: 10.3389/fphys.2023.1182505.