[On the Stress State of Bone Cement Fixing a Hip Joint Prosthesis (author's Transl)]
Overview
Overview
Journal
Arch Orthop Unfallchir
Specialties
Emergency Medicine
Orthopedics
Orthopedics
Date
1977 Jun 30
PMID
889525
Citations
1
Authors
Authors
Affiliations
Affiliations
Soon will be listed here.
Abstract
Using a simplified model, tests were made to investigate the shear stress distribution at a hip joint prosthesis. A marked increase of stress at the tip of the prosthesis was found through photoelasticity. This peak stress resulted in a safety factor of 3 against exceeding the shear strength of bone cement. Since the calculation of the stress is based on unfavourable assumptions, the actual factor of safety should show a higher value still.