The Effect of a Portable HEPA-filtered Body Exhaust System on Airborne Microbial Contamination in a Conventional Operating Room
Overview
Infectious Diseases
Nursing
Public Health
Authors
Affiliations
Objective: To study the effect of a portable HEPA-filtered air exhaust system (Stackhouse Freedom Surgical Helmet System) on airborne microbial contamination in a modern conventional operating room.
Design And Setting: Microbial air sampling was done with a two-stage Anderson sampler at the wound site during 46 total joint replacements. All operations were performed by the same surgeon in the same operating room at a large community hospital.
Results: In 18 cases done without air exhaust hoods, the number of bacterial and fungal colony-forming units (CFU) ranged from 0.6 to 11.7 (mean, 3.6). Air sampling during 28 operations with the operating team in air exhaust hoods revealed a mean of 3.6 CFU (range, 0 to 11.4). Bacterial CFU averaged 3.4 without hoods and 3.2 with exhaust hoods. Coagulase-negative staphylococci were the most common isolates (48% of isolates with hood, 55% without hood). No infections occurred.
Conclusion: We concluded that these air exhaust hoods did not lower airborne microbial contamination detectable with this air sampling method, as compared to standard head cover and mask, in a modern conventional operating room.
Personal Protective Equipment and COVID-19: A Review for Surgeons.
Stewart C, Thornblade L, Diamond D, Fong Y, Melstrom L Ann Surg. 2020; 272(2):e132-e138.
PMID: 32675516 PMC: 7268841. DOI: 10.1097/SLA.0000000000003991.
Surgeon Personal Protection: An Underappreciated Benefit of Positive-pressure Exhaust Suits.
Makovicka J, Bingham J, Patel K, Young S, Beauchamp C, Spangehl M Clin Orthop Relat Res. 2018; 476(6):1341-1348.
PMID: 29481350 PMC: 6263565. DOI: 10.1007/s11999.0000000000000253.
Nakajima D, Tateiwa T, Masaoka T, Takahashi Y, Shishido T, Yamamoto K Eur J Orthop Surg Traumatol. 2016; 27(8):1139-1143.
PMID: 27796583 DOI: 10.1007/s00590-016-1874-8.
Contamination Relative to the Activation Timing of Filtered-Exhaust Helmets.
Hanselman A, Montague M, Murphy T, Dietz M J Arthroplasty. 2015; 31(4):776-80.
PMID: 26684270 PMC: 5513156. DOI: 10.1016/j.arth.2015.10.039.
The Gown-glove Interface Is a Source of Contamination: A Comparative Study.
Fraser J, Young S, Valentine K, Probst N, Spangehl M Clin Orthop Relat Res. 2014; 473(7):2291-7.
PMID: 25488405 PMC: 4457760. DOI: 10.1007/s11999-014-4094-8.