Insights from Three-dimensional Echocardiographic Laser Stereolithography. Effect of Leaflet Funnel Geometry on the Coefficient of Orifice Contraction, Pressure Loss, and the Gorlin Formula in Mitral Stenosis

Overview
Journal Circulation
Date 1996 Aug 1
PMID 8759088
Citations 8
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Three-dimensional echocardiography can allow us to address uniquely three-dimensional scientific questions, for example, the hypothesis that the impact of a stenotic valve depends not only on its limiting orifice area but also on its three-dimensional geometry proximal to the orifice. This can affect the coefficient of orifice contraction (Cc = effective/anatomic area), which is important because for a given flow rate and anatomic area, a lower Cc gives a higher velocity and pressure gradient, and Cc, routinely assumed constant in the Gorlin equation, may vary with valve shape (60% for a flat plate, 100% for a tube). To date, it has not been possible to study this with actual valve shapes in patients.

Methods And Results: Three-dimensional echocardiography reconstructed valve geometries typical of the spectrum in patients with mitral stenosis: mobile doming, intermediate conical, and relatively flat immobile valves. Each geometry was constructed with orifice areas of 0.5, 1.0 and 1.5 cm2 by stereolithography (computerized laser polymerization) (total, nine valves) and studied at physiological flow rates. Cc varied prominently with shape and was larger for the longer, tapered dome (more gradual flow convergence proximal and distal to the limiting orifice): for an anatomic orifice of 1.5 cm2, Cc increased from 0.73 (flat) to 0.87 (dome), and for an area of 0.5 cm2, from 0.62 to 0.75. For each shape, Cc increased with increasing orifice size relative to the proximal funnel (more tubelike). These variations translated into important differences of up to 40% in pressure gradient for the same anatomic area and flow rate (greatest for the flattest valves), with a corresponding variation in calculated Gorlin area (an effective area) relative to anatomic values.

Conclusions: The coefficient of contraction and the related net pressure loss are importantly affected by the variations in leaflet geometry seen in patients with mitral stenosis. Three-dimensional echocardiography and stereolithography, with the use of actual information from patients, can address such uniquely three-dimensional questions to provide insight into the relations between cardiac structure, pressure, and flows.

Citing Articles

Degenerative Mitral Stenosis-Diagnostic Challenges and Future Directions.

Joury A, Puleo C, Golive A, Gilliland Y, Pressman G, Qamruddin S Rev Cardiovasc Med. 2024; 23(10):354.

PMID: 39077129 PMC: 11267384. DOI: 10.31083/j.rcm2310354.


Echocardiographic diagnosis of adult valvular heart disease.

J Med Ultrason (2001). 2016; 43(4):557-97.

PMID: 27038348 DOI: 10.1007/s10396-016-0702-4.


Mitral valve area during exercise after restrictive mitral valve annuloplasty: importance of diastolic anterior leaflet tethering.

Bertrand P, Verbrugge F, Verhaert D, Smeets C, Grieten L, Mullens W J Am Coll Cardiol. 2015; 65(5):452-61.

PMID: 25660923 PMC: 4372048. DOI: 10.1016/j.jacc.2014.11.037.


Functional mitral stenosis after surgical annuloplasty for ischemic mitral regurgitation: importance of subvalvular tethering in the mechanism and dynamic deterioration during exertion.

Kubota K, Otsuji Y, Ueno T, Koriyama C, Levine R, Sakata R J Thorac Cardiovasc Surg. 2010; 140(3):617-23.

PMID: 20122701 PMC: 2972671. DOI: 10.1016/j.jtcvs.2009.11.003.


Assessing mitral valve area and orifice geometry in calcific mitral stenosis: a new solution by real-time three-dimensional echocardiography.

Chu J, Levine R, Chua S, Poh K, Morris E, Hua L J Am Soc Echocardiogr. 2008; 21(9):1006-9.

PMID: 18620839 PMC: 2980787. DOI: 10.1016/j.echo.2008.05.010.