Axial and Lateral Resolution of Rotational Intravascular Ultrasound: in Vitro Observations and Diagnostic Implications

Overview
Date 1995 Jul 1
PMID 8581904
Citations 2
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To determine the axial and lateral resolution of a rotating intravascular ultrasound system and the diagnostic implications for the diagnosis of early artherosclerosis.

Methods: The resolution of a 20 MHz rotating transducer was tested in a specially designed high-resolution phantom and in five aortic autopsy specimens with varying degrees of early atherosclerosis.

Results: The best lateral resolution is at the minimal distance between transducer and object. Measurements in the wire phantom showed this to be better than 0.43 mm. This is less than the axial resolution which is better than 0.13 mm. Decreasing lateral resolution with increasing distance between transducer and object is manifested by arcing and overestimation of the extent of focal atherosclerotic lesions.

Conclusion: Axial resolution is significantly better than lateral resolution. Rapid deterioration of lateral resolution affects the diagnostic ability to characterize the extent of early atherosclerosis. Eccentric positioning of the transducer tip, particularly in larger vessels, will therefore influence diagnostic accuracy in vivo.

Citing Articles

Safety of embolic protection device-assisted and unprotected intravascular ultrasound in evaluating carotid artery atherosclerotic lesions.

Musialek P, Pieniazek P, Tracz W, Tekieli L, Przewlocki T, Kablak-Ziembicka A Med Sci Monit. 2012; 18(2):MT7-18.

PMID: 22293887 PMC: 3560589. DOI: 10.12659/msm.882452.


A novel realistic three-layer phantom for intravascular ultrasound imaging.

Brunette J, Mongrain R, Cloutier G, Bertrand M, Bertrand O, Tardif J Int J Cardiovasc Imaging. 2002; 17(5):371-81.

PMID: 12025951 DOI: 10.1023/a:1011996415966.

References
1.
Yock P, Fitzgerald P, Sudhir K, Linker D, White W, Ports A . Intravascular ultrasound imaging for guidance of atherectomy and other plaque removal techniques. Int J Card Imaging. 1991; 6(3-4):179-89. DOI: 10.1007/BF01797850. View

2.
Meyer C, Chiang E, Fechner K, Fitting D, Williams D, Buda A . Feasibility of high-resolution, intravascular ultrasonic imaging catheters. Radiology. 1988; 168(1):113-6. DOI: 10.1148/radiology.168.1.3289084. View

3.
ten Hoff H, Korbijn A, Smith T, Klinkhamer J, Bom N . Imaging artifacts in mechanically driven ultrasound catheters. Int J Card Imaging. 1989; 4(2-4):195-9. DOI: 10.1007/BF01745150. View

4.
Gussenhoven W, Essed C, Frietman P, Mastik F, Lancee C, Slager C . Intravascular echographic assessment of vessel wall characteristics: a correlation with histology. Int J Card Imaging. 1989; 4(2-4):105-16. DOI: 10.1007/BF01745140. View

5.
Engeler C, Yedlicka J, Letourneau J, Hunter D, AMPLATZ K . 1990 ARRS Executive Council Award. Intravascular sonography in the detection of arteriosclerosis and evaluation of vascular interventional procedures. AJR Am J Roentgenol. 1991; 156(5):1087-90. DOI: 10.2214/ajr.156.5.1826804. View