A trial of the Nellcor Stat Cap, which detects exhaled carbon dioxide, as an aid to determining whether endotracheal tubes are placed in the oesophagus or the trachea, was carried out in the neonatal unit of this hospital. Twenty five neonates, with a mean (range) gestational age of 33 (24-42) weeks and a mean (range) birthweight of 2.17 (0.54-4.1) kg were studied over two months. These babies underwent a total of 58 intubations and 20 suspected self-extubations. The Nellcor Stat Cap was easy to use. It confirmed clinical findings on 20/20 occasions when the endotracheal tube was in the oesophagus and on 57/58 (98%) occasions when it was in the trachea. On 14/78 (19%) occasions the machine provided helpful additional information in reaching a decision on the adequacy of tube placement. Failure to detect rapidly accidental oesophageal intubation or unintentional tracheal extubation can result in rapid deterioration of the condition of ventilated newborns. The machine is a valuable aid to intubation and rapid diagnosis of self-extubation.

Citing Articles

Question 2. Should carbon dioxide detectors be used to check correct placement of endotracheal tubes in preterm and term neonates?.

Gowda H Arch Dis Child. 2011; 96(12):1201-3.

PMID: 22080462 PMC: 3223393. DOI: 10.1136/archdischild-2011-300893.

References
1.
KALENDA Z . The capnogram as a guide to the efficacy of cardiac massage. Resuscitation. 1978; 6(4):259-63. DOI: 10.1016/0300-9572(78)90006-0. View

2.
Pollard B, Junius F . Accidental intubation of the oesophagus. Anaesth Intensive Care. 1980; 8(2):183-6. DOI: 10.1177/0310057X8000800215. View

3.
HOWELLS T, Riethmuller R . Signs of endotracheal intubation. Anaesthesia. 1980; 35(10):984-6. DOI: 10.1111/j.1365-2044.1980.tb04997.x. View

4.
Linko K, Paloheimo M, TAMMISTO T . Capnography for detection of accidental oesophageal intubation. Acta Anaesthesiol Scand. 1983; 27(3):199-202. DOI: 10.1111/j.1399-6576.1983.tb01934.x. View

5.
MURRAY I, Modell J . Early detection of endotracheal tube accidents by monitoring carbon dioxide concentration in respiratory gas. Anesthesiology. 1983; 59(4):344-6. DOI: 10.1097/00000542-198310000-00013. View