» Articles » PMID: 8491011

Ultrasound-assisted Cannulation of the Internal Jugular Vein. A Prospective Comparison to the External Landmark-guided Technique

Overview
Journal Circulation
Date 1993 May 1
PMID 8491011
Citations 93
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Central venous access is an essential part of patient management in many clinical settings and is usually achieved with a blinded, external landmark-guided technique. The purpose of this study is to evaluate whether an ultrasound technique can improve on the traditional method.

Methods And Results: We prospectively evaluated an ultrasound-guided method in 302 patients undergoing internal jugular venous cannulation and compared the results with 302 patients in whom an external landmark-guided technique was used. Ultrasound was used exclusively in an additional 626 patients. Cannulation of the internal jugular vein was achieved in all patients (100%) using ultrasound and in 266 patients (88.1%) using the landmark-guided technique (p < 0.001). The vein was entered on the first attempt in 78% of patients using ultrasound and in 38% using the landmark technique (p < 0.001). Average access time (skin to vein) was 9.8 seconds (2-68 seconds) by the ultrasound approach and 44.5 seconds (2-1,000 seconds) by the landmark approach (p < 0.001). Using ultrasound, puncture of the carotid artery occurred in 1.7% of patients, brachial plexus irritation in 0.4%, and hematoma in 0.2%. In the external landmark group, puncture of the carotid artery occurred in 8.3% of patients (p < 0.001), brachial plexus irritation in 1.7% (p < 0.001), and hematoma in 3.3% (p < 0.001).

Conclusions: Ultrasound-guided cannulation of the internal jugular vein significantly improves success rate, decreases access time, and reduces complication rate. These results suggest that this technique may be preferred in complicated cases or when access problems are anticipated.

Citing Articles

A novel "lateral approach short axis in-plane" technique vs. conventional "short-axis out-of-plane approach" for ultrasound-guided internal jugular vein access: a prospective randomized non-inferiority trial.

Kalina M, Vargova P, Bubenikova A, Skulec R, Cerny V, Astapenko D Ultrasound J. 2025; 17(1):5.

PMID: 39821817 PMC: 11739437. DOI: 10.1186/s13089-025-00405-9.


Stepping into the Virtual Realm: Examining the Utility of Non-Immersive VR on Medical Residents' Self-Efficacy, Performance, and Satisfaction.

Gonzalez-Vargas J, Sinz E, Moore J, Miller S Hum Factors Health. 2024; 5.

PMID: 39431120 PMC: 11486498. DOI: 10.1016/j.hfh.2024.100072.


Non-tunneled haemodialysis catheter-related blood stream infections and associated factors among first time haemodialysis patients: a prospective study from a tertiary care hospital in Sri Lanka.

Muthukuda C, Suriyakumara V, Samarathunga T, Liyanage L, Marasinghe A BMC Nephrol. 2024; 25(1):280.

PMID: 39198819 PMC: 11361233. DOI: 10.1186/s12882-024-03726-4.


Are routine chest radiographs still indicated after central line insertion? A scoping review.

Brindley P, Deschamps J, Milovanovic L, Buchanan B J Intensive Care Soc. 2024; 25(2):190-207.

PMID: 38737308 PMC: 11086721. DOI: 10.1177/17511437241227739.


Internal thoracic vein cannulation for venous port insertion.

Berczi A, Osztrogonacz P, Csobay-Novak C J Vasc Surg Venous Lymphat Disord. 2024; 13(1):101887.

PMID: 38570163 PMC: 11764252. DOI: 10.1016/j.jvsv.2024.101887.