Evaluation of Various Models for Respiratory Inductance Plethysmography Calibration

Overview
Date 1993 Mar 1
PMID 8482659
Citations 11
Authors
Affiliations
Soon will be listed here.
Abstract

We evaluated one nonlinear and two linear models of the ventilatory system while calibrating the respiratory inductance plethysmograph (RIP) against a pneumotachometer. A calibration method involving voluntary varying rib cage and abdominal contributions to tidal volume in a single body position was utilized. The influence on accuracy of the choice of respiratory phase during calibration was assessed. Both tidal and intratidal volumes were evaluated. Ten adults with no history of respiratory disorders went through RIP calibration and validation in the sitting and supine positions. A linear calibration model, relating lung volume changes from the start of inspiration or expiration to rib cage and abdominal excursions from initiation of respiratory motion, had the best accuracy. The choice of respiratory phase for calibration did not affect accuracy. RIP generally underestimated lung volume at the start of inspiration and overestimated lung volume at the end of inspiration. RIP was more accurate in the supine than the sitting position, probably because of limited spine flexion in the supine position.

Citing Articles

Comparison of Respiratory Calibration Methods for the Estimation of Lung Volume in Children With and Without Neuromotor Disorders.

Darling-White M J Speech Lang Hear Res. 2022; 65(2):525-537.

PMID: 35050707 PMC: 9132159. DOI: 10.1044/2021_JSLHR-21-00333.


Telemonitoring Techniques for Lung Volume Measurement: Accuracy, Artifacts and Effort.

Mannee D, de Jongh F, van Helvoort H Front Digit Health. 2021; 2:559483.

PMID: 34713036 PMC: 8521879. DOI: 10.3389/fdgth.2020.559483.


End-expiratory lung volume remains stable during N MBW in healthy sleeping infants.

Gustafsson P, Kadar L, Kjellberg S, Andersson L, Lindblad A, Robinson P Physiol Rep. 2020; 8(16):e14477.

PMID: 32845567 PMC: 7448799. DOI: 10.14814/phy2.14477.


The Accuracy of Respiratory Calibration Methods for Estimating Lung Volume During Speech Breathing: A Comparison of Four Methods Across Three Adult Cohorts.

McKenna V, Huber J J Speech Lang Hear Res. 2019; 62(8):2632-2644.

PMID: 31330112 PMC: 6802911. DOI: 10.1044/2019_JSLHR-S-18-0478.


Respiratory modulation of startle eye blink: a new approach to assess afferent signals from the respiratory system.

Schulz A, Schilling T, Vogele C, Larra M, Schachinger H Philos Trans R Soc Lond B Biol Sci. 2017; 371(1708).

PMID: 28080976 PMC: 5062107. DOI: 10.1098/rstb.2016.0019.