Abstract

The frequency of ocular microtremor (OMT) is related to the functional status of the brain stem, and thus OMT may be useful in the diagnosis and management of brain stem disorders. The paper discusses the design of an OMT measurement system and reports quantitative specifications for three portable systems. All systems use a piezo-electric element as the transducer, which measures the displacement of the sclera during eye rotations. The systems differ in the manner in which the signal is recorded. All systems can detect eye movements corresponding to displacements of the sclera ranging from 12 to over 3000 nm. The frequency responses of all systems are flat (< 2 dB deviation from peak response) between 20 and 150 Hz. The phase response shows deviations (< pi) at the extremes of this range, but qualitative comparison of input and measured signals demonstrates that phase distortion is not excessive. Thus all systems are acceptable for clinical studies involving OMT.

Citing Articles

Digital measurement of ocular microtremor in Parkinson's disease: Protocol for a pilot study to assess reliability and clinical validation.

Graham L, Vitorio R, Walker R, Godfrey A, Morris R, Stuart S PLoS One. 2025; 20(1):e0313452.

PMID: 39775540 PMC: 11709268. DOI: 10.1371/journal.pone.0313452.


A Bayesian computational model for online character recognition and disability assessment during cursive eye writing.

Diard J, Rynik V, Lorenceau J Front Psychol. 2013; 4:843.

PMID: 24273525 PMC: 3822325. DOI: 10.3389/fpsyg.2013.00843.


Simultaneous recordings of ocular microtremor and microsaccades with a piezoelectric sensor and a video-oculography system.

McCamy M, Collins N, Otero-Millan J, Al-Kalbani M, Macknik S, Coakley D PeerJ. 2013; 1:e14.

PMID: 23638348 PMC: 3629042. DOI: 10.7717/peerj.14.

References
1.
Eizenman M, Frecker R, HALLETT P . Precise non-contacting measurement of eye movements using the corneal reflex. Vision Res. 1984; 24(2):167-74. DOI: 10.1016/0042-6989(84)90103-2. View

2.
Abakumova L, Shakhnovich A, Thomas J . An investigation of the correlation between abnormal patterns of ocular microtremor and an abnormal pupil reflex in neurological patients. J Neurol Sci. 1975; 26(4):469-77. DOI: 10.1016/0022-510x(75)90047-7. View

3.
Michalik M . [Ocular microtremor--an indicator of the functional status of the brain stem]. Psychiatr Neurol Med Psychol Beih. 1983; 29:196-204. View

4.
Michalik M . [Spectral analyses in ocular microtremor in disorders of brain stem function]. EEG EMG Z Elektroenzephalogr Elektromyogr Verwandte Geb. 1987; 18(1):20-6. View

5.
Coakley D, Thomas J . The ocular microtremor record as a potential procedure for establishing brain death. J Neurol Sci. 1977; 31(2):199-205. DOI: 10.1016/0022-510x(77)90106-x. View