Three-dimensional Treatment Planning of Intracavitary Gynecologic Implants: Analysis of Ten Cases and Implications for Dose Specification

Overview
Specialties Oncology
Radiology
Date 1994 Jan 1
PMID 8270452
Citations 17
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: Results of 3-dimensional treatment planning for ten intracavitary gynecologic implants and implications for dose specification are presented.

Methods And Materials: Using a computed tomographic (CT) compatible intracavitary applicator we have performed CT scans during gynecologic brachytherapy in 10 cases. A CT-based treatment planning system with 3-dimensional capabilities was used to calculate and display dose in three dimensions. Conventional point doses including the estimated bladder and rectal maximum doses and dose to Point A were acquired from orthogonal simulation films. CT maximum bladder and rectal doses and minimum cervix doses were ascertained from isodose lines displayed on individual CT images. Dose volume histograms for the bladder, rectum and cervix were generated and used to obtain volume of the cervix target volume receiving less than the prescribed dose and the volume of bladder and rectum receiving more than the orthogonal maximum doses. The 5 cc volume of bladder and rectum receiving the highest dose were also calculated.

Results: Average values of CT point doses and volumes are compared with the traditionally obtained doses. As demonstrated by others, much higher bladder and rectal doses are found using the CT information. The minimum dose to the cervix target volume is lower than the dose to Point A in each case. CT maximum bladder and rectum and minimum cervix target doses may not be the best index doses to correlate with outcome because of the small volumes receiving the dose.

Conclusion: We hypothesize that clinically useful bladder, rectal and cervix target volume doses will include volume information which is obtainable with dose volume histogram analysis.

Citing Articles

Dosimetric analysis of intracavitary brachytherapy applicators: a practical study.

Khan A, Vijay A, Wani S, Haq M Radiat Oncol J. 2022; 40(3):180-191.

PMID: 36200307 PMC: 9535410. DOI: 10.3857/roj.2022.00199.


Description of a novel technique for ultrasound-based planning for gynaecological 3D brachytherapy and comparison between plans of this technique and 2D with fluoroscopy.

Salas J, Veliz D Ecancermedicalscience. 2022; 16:1415.

PMID: 36072239 PMC: 9377822. DOI: 10.3332/ecancer.2022.1415.


Dose Summation Strategies for External Beam Radiation Therapy and Brachytherapy in Gynecologic Malignancy: A Review from the NRG Oncology and NCTN Medical Physics Subcommittees.

Kim H, Lee Y, Benedict S, Dyer B, Price M, Rong Y Int J Radiat Oncol Biol Phys. 2021; 111(4):999-1010.

PMID: 34147581 PMC: 8594937. DOI: 10.1016/j.ijrobp.2021.06.019.


Evaluation of variation of interfraction doses to organs at risk during brachytherapy of cervical cancer.

Mukundan H, Tyagi K, Mukherjee D, Patel R Med J Armed Forces India. 2020; 76(2):201-206.

PMID: 32476719 PMC: 7244865. DOI: 10.1016/j.mjafi.2019.02.004.


Evaluation of Volumetric Doses of Organs at Risk in Carcinoma Cervix Patients with HDR Intracavitary Brachytherapy and Comparison of CT-based and Conventional Plans.

Srivastava S, Painuly N, Mishra S, Srivastava K, Singh N, Singh S J Biomed Phys Eng. 2020; 9(6):603-612.

PMID: 32039090 PMC: 6943852. DOI: 10.31661/jbpe.v0i0.867.