The Effect of Pulsating Arteries on Reflectance Pulse Oximetry: Measurements in Adults and Neonates

Overview
Journal J Clin Monit
Specialty Biotechnology
Date 1995 Mar 1
PMID 7760084
Citations 4
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: The objective of our study was to describe the results from human experiments during normoxia that demonstrate the effect of pulsating arteries on the measured arterial oxygen saturation (SpO2) using a reflectance pulse oximeter sensor.

Methods: In 6 healthy adults and 7 healthy neonates, a Nellcor reflection sensor (FS-10 oxisensor, Nellcor, Inc., Pleasanton, CA) was placed in three different positions: (1) on the forehead, (2) on the temporal area, with the photodiode placed over the superficial temporal artery, and (3) on the temporal area, with the light-emitting diodes (LEDs) placed over the superficial temporal artery.

Results: Placement of the sensor in position 2 resulted in a significantly lower SpO2 reading, compared to sensor position 1: 5.8% (p < 0.01) lower for adults and 7.5% (p < 0.01) lower for neonates. Placement of the sensor in position 3 resulted in significantly larger plethysmographic signals, compared to sensor position 1; but, the Spo2 readings were alike.

Conclusions: Pulsating arteries can affect the reliability of reflection pulse oximetry. Depending on the position of the sensor, a falsely low Spo2 value can be observed.

Citing Articles

The pulse in reflectance pulse oximetry: modeling and experimental studies.

Reuss J, Siker D J Clin Monit Comput. 2005; 18(4):289-99.

PMID: 15779841 DOI: 10.1007/s10877-005-2909-6.


The influence of larger subcutaneous blood vessels on pulse oximetry.

Mannheimer P, O N, Konecny E J Clin Monit Comput. 2004; 18(3):179-88.

PMID: 15562984 DOI: 10.1023/b:jocm.0000042924.16052.5e.


Factors influencing fetal pulse oximetry performance.

Reuss J J Clin Monit Comput. 2004; 18(1):13-24.

PMID: 15139579 DOI: 10.1023/b:jocm.0000025278.82852.b3.


Reflectance pulse oximetry in fetal lambs: subcutaneous vessels and vasoconstriction affect its reliability.

Nijland R, Jongsma H, Verbruggen I, Nijhuis J J Clin Monit. 1996; 12(3):225-30.

PMID: 8823646 DOI: 10.1007/BF00857643.

References
1.
Mendelson Y, Ochs B . Noninvasive pulse oximetry utilizing skin reflectance photoplethysmography. IEEE Trans Biomed Eng. 1988; 35(10):798-805. DOI: 10.1109/10.7286. View

2.
Severinghaus J, Naifeh K, Koh S . Errors in 14 pulse oximeters during profound hypoxia. J Clin Monit. 1989; 5(2):72-81. DOI: 10.1007/BF01617877. View

3.
Cheng E, Hopwood M, Kay J . Forehead pulse oximetry compared with finger pulse oximetry and arterial blood gas measurement. J Clin Monit. 1988; 4(3):223-6. DOI: 10.1007/BF01621821. View

4.
Dildy G, Clark S, Loucks C . Preliminary experience with intrapartum fetal pulse oximetry in humans. Obstet Gynecol. 1993; 81(4):630-5. View

5.
de Kock J, Tarassenko L . In vitro investigation of the factors affecting pulse oximetry. J Biomed Eng. 1991; 13(1):61-6. DOI: 10.1016/0141-5425(91)90046-a. View