Study Design: The biomechanics of three different instrumentation constructs applied at the destabilized cervicothoracic junction were evaluated.

Objectives: To find an efficient way in restoring stability of the cervicothoracic junction in cases with and without laminectomy.

Summary Of Background Data: Different instrumentation techniques have been evaluated biomechanically and used clinically for managing instabilities between the fourth and sixth cervical vertebrae. These constructs have not been evaluated at the cervicothoracic junction.

Methods: Six human spines were tested nondestructively in axial torsion, flexion, and extension with the C6-T2 motion segments left unconstrained. The three-dimensional displacements and rotations between C7 and T1 vertebrae were measured using a sonic digitizer. After intact testing, a distractive-flexion Stage 3 cervical spinal injury was simulated surgically between C7 and T1. The specimens underwent sequential instrumentation and mechanical testing with three constructs: posterior Synthes lateral mass plate, posterior pediatric Cotrel-Dubousset rod system with lamina hooks and a crosslink, and anterior Synthes cervical locking plate.

Results: Posterior stabilization techniques had statistically more stiffness than anterior plates. The Cotrel-Dubousset system offered the largest stiffness ratio (instrumented/intact) in flexion, extension, and rotation. There was no statistical difference between posterior plates and Cotrel-Dubousset instrumentation. The stiffness of the anterior plate did not differ significantly from the intact spine.

Conclusion: Our data show that instability of the cervicothoracic junction can be efficiently restored by either anterior plates, posterior plates, or posterior hook-rod constructs (Cotrel-Dubousset). Posterior constructs showed increased stiffness over anterior plates.

Citing Articles

The Efficacy of Cervical Pedicle Screw Is Enhanced When Used With 5.5-mm Rods for Metastatic Cervical Spinal Tumor Surgery.

Park D, Lee S, Lee S, Park J, Yang H, Kim C Neurospine. 2024; 21(1):352-360.

PMID: 38291748 PMC: 10992656. DOI: 10.14245/ns.2346778.389.


Clinical and radiographic benefits of skipping C7 instrumentation in posterior cervicothoracic fusion: a retrospective analysis.

Patel S, Sadeh M, Tobin M, Chaudhry N, Gragnaniello C, Neckrysh S J Spine Surg. 2022; 8(3):333-342.

PMID: 36285098 PMC: 9547698. DOI: 10.21037/jss-21-85.


Traumatic Fracture: Dislocation of Cervicothoracic Junction-Grand Round Presentation of C7-T1 Instabilities and Different Instrumentation Techniques.

Alsofyani M, Ghailane S, Alsalmi S, Jakinapally S, Boissiere L, Obeid I Case Rep Orthop. 2020; 2020:7578628.

PMID: 32665871 PMC: 7349622. DOI: 10.1155/2020/7578628.


Cortical Bone Trajectory Screws for Fixation Across the Cervicothoracic Junction: Surgical Technique and Outcomes.

Obeidat M, Tan Z, Finkelstein J Global Spine J. 2019; 9(8):859-865.

PMID: 31819852 PMC: 6882086. DOI: 10.1177/2192568219838822.


C7 intra-laminar screws for complex cervicothoracic spine surgery-a case series.

Wilson P, Selby M J Spine Surg. 2018; 3(4):561-566.

PMID: 29354732 PMC: 5760425. DOI: 10.21037/jss.2017.10.05.