» Articles » PMID: 737092

Effects of Recording Speed on Precision of Time-based Polycardiographic Measurements. Optimal Paper Speeds for Measuring Points and Intervals

Overview
Journal Br Heart J
Date 1978 Dec 1
PMID 737092
Citations 5
Authors
Affiliations
Soon will be listed here.
Abstract

Optimal paper speeds have not been established for all time-based measurements of the cardiac cycle by appropriately designed observer performance studies. In 10 subjects (5 normals and 5 cardiac patients) carotid pulse, phonocardiogram, and electrocardiogram were recorded on magnetic tape for measurement of all fiducial points for systolic time intervals, the systolic time intervals themselves, the pulse transmission time, cycle length (RR), qR time, and R-to-point versus q-to-point measurements at recording speeds of 25, 50, 75, 100, and 200 mm/s. Tracings were coded numerically and randomised. Three observers measured all points and calculated intervals in a sequence determined by individual tables of random numbers. Left ventricular ejection time was the only calculation that could be made at 25 mm/s statistically equally well as at all other speeds. The smallest numerical observer differences occurred uniformly at 100 mm/s paper speed when all recording speeds were considered. However, after excluding the 25 mm/s speed there were no significant differences among point measurements. Measurements of points from R (rather than q) reduced observer variability. We conclude that for point measurements, for systolic time intervals, pulse transmission time, and RR interval, recording speed between 50 and 200 mm/s showed no statistical differences, though smallest numerical differences occurred at 100 mm/s. For LVET, 25 mm/s was satisfactory.

Citing Articles

Systolic time intervals in clinical pharmacology.

Li Q, Belz G Eur J Clin Pharmacol. 1993; 44(5):415-21.

PMID: 8359176 DOI: 10.1007/BF00315536.


Isovolumic contraction time of right ventricle in d-transposition of great arteries.

Fouron J, Vallot F, Bourlon F, Lombaert M, Ducharme G, DAVIGNON A Br Heart J. 1980; 44(2):204-7.

PMID: 7426174 PMC: 482383. DOI: 10.1136/hrt.44.2.204.


Mechanocardiographic assessment of systolic time intervals in normal children.

Ulmer H, Heupel E, Weckesser G Basic Res Cardiol. 1982; 77(2):197-212.

PMID: 7092778 DOI: 10.1007/BF01908173.


Systolic time intervals: a review of the method in the non-invasive investigation of cardiac function in health, disease and clinical pharmacology.

Hassan S, Turner P Postgrad Med J. 1983; 59(693):423-34.

PMID: 6353394 PMC: 2417541. DOI: 10.1136/pgmj.59.693.423.


Cardiovascular effects of static and dynamic exercise.

Chapman J, ELLIOTT P Eur J Appl Physiol Occup Physiol. 1988; 58(1-2):152-7.

PMID: 3203661 DOI: 10.1007/BF00636619.

References
1.
Lewis R, Rittogers S, Froester W, Boudoulas H . A critical review of the systolic time intervals. Circulation. 1977; 56(2):146-58. DOI: 10.1161/01.cir.56.2.146. View

2.
Pigott V, Elber E, Spodick D . The influence of recording speed on apexcardiographic timing. A multi-observer study of precision and performance utilizing randomized tracings in multiple subjects. Cardiology. 1972; 57(4):232-9. DOI: 10.1159/000169522. View

3.
Spodick D, Kumar S, Flessas A, Sriratanaban D, Sosler G, Klints R . Relationship of paper speed to precision of pulse wave measurements: a multiple observer study of left ventricular ejection time. Aerosp Med. 1969; 40(7):707-9. View