» Articles » PMID: 6983535

Quantitative Measurement of Bone Density Using Gamma-ray Computed Tomography

Overview
Specialty Radiology
Date 1982 Dec 1
PMID 6983535
Citations 8
Authors
Affiliations
Soon will be listed here.
Abstract

A special purpose gamma-ray computed tomography scanner has been developed for precise measurements of bone density in the human appendicular skeleton. Details of the scanner's hardware and of the software organization for system control and data analysis are given, together with an outline of the theoretical basis for conversion of measured linear attenuation coefficients to physical bone densities. Performance of the system was evaluated on bone-like phantoms. Clinically, a precision of +/- 0.5% is obtained for bone density determinations. This device is being used in experimental studies and clinical investigations.

Citing Articles

Quantitative CT reference values for vertebral trabecular bone density in children and young adults.

Gilsanz V, Perez F, Campbell P, Dorey F, Lee D, Wren T Radiology. 2008; 250(1):222-7.

PMID: 19001149 PMC: 2817831. DOI: 10.1148/radiol.2493080206.


Evolution of spinal bone loss and biochemical markers of bone remodeling after menopause in normal women.

Reeve J, Pearson J, Mitchell A, Green J, Nicholls A, Justice J Calcif Tissue Int. 1995; 57(2):105-10.

PMID: 7584869 DOI: 10.1007/BF00298429.


Restoring lost bone in osteoporosis.

Lee R CMAJ. 1985; 133(9):847-50.

PMID: 4052896 PMC: 1346292.


Skeletal challenge: an experimental study of pharmacologically induced changes in bone density in the distal radius, using gamma-ray computed tomography.

Hangartner T, Overton T, Harley C, van den Berg L, Crockford P Calcif Tissue Int. 1985; 37(1):19-24.

PMID: 3922592 DOI: 10.1007/BF02557673.


Quantitative computed tomography: update 1987.

Genant H, Steiger P, Block J, Glueer C, Ettinger B, Harris S Calcif Tissue Int. 1987; 41(4):179-86.

PMID: 3119173 DOI: 10.1007/BF02555236.