Sphygmomanometers in Hospital and Family Practice: Problems and Recommendations

Overview
Specialty General Medicine
Date 1982 Aug 14
PMID 6809128
Citations 30
Authors
Affiliations
Soon will be listed here.
Abstract

The accuracy and working condition of 210 sphygmomanometers were tested: 100 (50 and mercury and 50 aneroid) models were used in family practices and 100 mercury models in hospitals. Faults in the inflation-deflation system were common and caused mainly by dirt or wear in the control valves. Leakage occurred in 48% of the hospital and 33% of the family practice sphygmomanometers. In the mercury models the mercury or air vents were often in an unsatisfactory condition or the calibrated glass tube dirty. The accuracy of the gauges was examined at 90 and 150 mm Hg: fewer than 2% of the mercury sphygmomanometers but 30% of the aneroid models had errors greater than +/- 4 mm Hg at either pressure. Over half of the cuffs examined had bladders widths less than the recommended size, and 94% had bladders shorter than the length recommended for use on normal adults. Mercury sphygmomanometers should be bought in preference to aneroid models as they are more accurate, less expensive in the long term, and can be maintained by the owner; they should be checked every six to 12 months depending on usage. Replacement parts should be kept readily available.

Citing Articles

Aneroid Auscultatory Sphygmomanometers and Automated Oscillometric Devices as Mercury-Free Alternatives in Children.

Song Y Korean Circ J. 2024; 54(5):288-290.

PMID: 38767341 PMC: 11109841. DOI: 10.4070/kcj.2024.0115.


Blood pressure measurements and hypertension in infants, children, and adolescents: from the postmercury to mobile devices.

Lim S, Kim S Clin Exp Pediatr. 2021; 65(2):73-80.

PMID: 34530519 PMC: 8841968. DOI: 10.3345/cep.2021.00143.


Office blood pressure measurement: A comprehensive review.

Siddique S, Khan A, Shahab H, Zhang Y, Tay J, Buranakitjaroen P J Clin Hypertens (Greenwich). 2021; 23(3):440-449.

PMID: 33420745 PMC: 8029522. DOI: 10.1111/jch.14169.


Pediatric hypertension: an updated review.

Ashraf M, Irshad M, Parry N Clin Hypertens. 2020; 26(1):22.

PMID: 33292828 PMC: 7706222. DOI: 10.1186/s40885-020-00156-w.


Ambulatory Blood Pressure Monitoring in Pediatrics.

Patel S, Daniels S Curr Hypertens Rep. 2019; 21(9):71.

PMID: 31350605 DOI: 10.1007/s11906-019-0976-1.


References
1.
King G . Taking the blood pressure. JAMA. 1969; 209(12):1902-4. View

2.
Conceicao S, Ward M, Kerr D . Defects in sphygmomanometers: an important source of error in blood pressure recording. Br Med J. 1976; 1(6014):886-8. PMC: 1639553. DOI: 10.1136/bmj.1.6014.886. View

3.
Shaw A, Deehan C, Lenihan J . Sphygmomanometers: errors due to blocked vents. Br Med J. 1979; 1(6166):789-90. PMC: 1598462. DOI: 10.1136/bmj.1.6166.789. View

4.
OBrien E, OMalley K . The sphygmomanometer. Br Med J. 1979; 2(6194):851-3. PMC: 1596669. DOI: 10.1136/bmj.2.6194.851. View

5.
Bowman C . Blood pressure errors with aneroid sphygmomanometers. Lancet. 1981; 1(8227):1005. DOI: 10.1016/s0140-6736(81)91776-1. View