» Articles » PMID: 666553

[2nd Generaltion of the Munich Hip-joint-simulator (author's Transl)]

Overview
Specialty Orthopedics
Date 1978 May 30
PMID 666553
Citations 2
Authors
Affiliations
Soon will be listed here.
Abstract

In addition to the improved, universal unsable, electronic-hydraulic directed hip-joint-simulator type "MUNCHEN I", a new simulator with simplified construction has been developed. This mechanical hip-joint-simulator is marked by a dual drive of movement and strain mechanism, that has been realized by a toothed belt drive. The adjustable, three-axial movement simulates level walking. According to its special, easier design the new simulator is highly adapted for comparative testing of total hip endoprostheses regarding the fields of quality improvement or standardization testing.

Citing Articles

[Implant wear and aseptic loosening. An overview].

Kaddick C, Catelas I, Pennekamp P, Wimmer M Orthopade. 2009; 38(8):690-7.

PMID: 19657620 DOI: 10.1007/s00132-009-1431-9.


Technological test-methods for hip prostheses.

Ungethum M Arch Orthop Trauma Surg (1978). 1980; 97(4):285-92.

PMID: 7458613 DOI: 10.1007/BF00380710.

References
1.
Ungethum M, Hildebrandt J, Jager M, Mosle H . [A new simulating system for artificial hip joints (author's transl)]. Arch Orthop Unfallchir. 1973; 77(4):304-14. DOI: 10.1007/BF00418916. View

2.
Hinterberger J, Ungethum M . [Investigations on tribology and strength of aluminium-oxide-ceramic--hip endoprostheses (author's transl)]. Z Orthop Ihre Grenzgeb. 1978; 116(3):285-303. View