Evaluation of a Two-minute Test for Urine Screening

Overview
Specialty Microbiology
Date 1983 Sep 1
PMID 6355154
Citations 12
Authors
Affiliations
Soon will be listed here.
Abstract

A study was conducted to evaluate the ability of a urine filtration system (Bac-T-Screen, Marion Laboratories, Inc., Kansas City, Mo.) to detect negative urine cultures within 2 min. A total of 1,000 urine specimens were tested with the Bac-T-Screen and compared with a standard semiquantitative culture plate method and the Autobac system (General Diagnostics, Warner-Lambert Co., Morris Plains, N.J.). Of the 1,000 clean voided urine specimens tested, 246 specimens had colony counts greater than or equal to 10(5) CFU/ml by the culture plate method. Of these, the Bac-T-Screen detected 65.4% (161 of 246), and the Autobac detected 63.0% (155 of 246). When pure cultures of diphtheroids, lactobacilli, and viridans streptococci other than group D and cultures containing multiple organisms were considered to be contaminants and, therefore, were excluded, there were 106 pure cultures of probable pathogens of which the Bac-T-Screen detected 76.4% (81 of 106) and the Autobac detected 90.6% (96 of 106). Some 133 specimens were uninterpretable with the Bac-T-Screen because 36 clogged the filter and 97 left a residual pigment on the filter. A majority of those clogging the filter (69.4%) had positive plate counts, whereas the majority of the pigmented urines had negative plate counts. Of those urine specimens tested. 754 were negative by the culture plate method. The false-positive rates for Bac-T-Screen and Autobac were 16.2 and 5.8%, respectively. As a urine screen, the Bac-T-Screen has a negative predictive value comparable to the Autobac system and has the advantage of being a 2-min test.

Citing Articles

Controversies in the laboratory diagnosis of community-acquired urinary tract infection.

Morgan M, McKenzie H Eur J Clin Microbiol Infect Dis. 1993; 12(7):491-504.

PMID: 8404909 DOI: 10.1007/BF01970954.


Evaluation of two bioluminescence-measuring instruments, the Turner Design and Lumac systems, for the rapid screening of urine specimens.

Welch W, Thompson L, Layman M, Southern Jr P J Clin Microbiol. 1984; 20(6):1165-70.

PMID: 6394619 PMC: 271538. DOI: 10.1128/jcm.20.6.1165-1170.1984.


Reverse-flow apparatus for enhanced colorimetric detection of bacteriuria.

Wallis C, Melnick J J Clin Microbiol. 1984; 20(5):942-4.

PMID: 6392325 PMC: 271479. DOI: 10.1128/jcm.20.5.942-944.1984.


Clinical evaluation of three rapid methods for the detection of significant bacteriuria.

Bertram M, Bruckner D J Clin Microbiol. 1985; 22(1):62-7.

PMID: 4019744 PMC: 268322. DOI: 10.1128/jcm.22.1.62-67.1985.


An instrument for the immediate quantification of bacteria in potable waters.

Wallis C, Melnick J Appl Environ Microbiol. 1985; 49(5):1251-3.

PMID: 4004238 PMC: 238537. DOI: 10.1128/aem.49.5.1251-1253.1985.


References
1.
Lamb V, Dalton H, WILKINS J . Electrochemical method for the early detection of urinary-tract infections. Am J Clin Pathol. 1976; 66(1):91-5. DOI: 10.1093/ajcp/66.1.91. View

2.
Ransohoff D, Feinstein A . Problems of spectrum and bias in evaluating the efficacy of diagnostic tests. N Engl J Med. 1978; 299(17):926-30. DOI: 10.1056/NEJM197810262991705. View

3.
Nicholson D, KOEPKE J . The Automicrobic System for urines. J Clin Microbiol. 1979; 10(6):823-33. PMC: 273278. DOI: 10.1128/jcm.10.6.823-833.1979. View

4.
Jenkins R, Hale D, Matsen J . Rapid semiautomated screening and processing of urine specimens. J Clin Microbiol. 1980; 11(3):220-5. PMC: 273367. DOI: 10.1128/jcm.11.3.220-225.1980. View

5.
Hale D, Wright D, McKie J, ISENBERG H, Jenkins R, Matsen J . Rapid screening for bacteriuria by light scatter photometry (Autobac): a collaborative study. J Clin Microbiol. 1981; 13(1):147-50. PMC: 273739. DOI: 10.1128/jcm.13.1.147-150.1981. View