Dermatologic Radiotherapy and Breast Cancer. Dose Measurements and Risk Quantification

Overview
Specialty Dermatology
Date 1982 Jan 1
PMID 6219629
Authors
Affiliations
Soon will be listed here.
Abstract

This study was set up to provide quantitative data to evaluate unsubstantiated claims that improper dermatologic radiation techniques may cause breast cancer. A thin mylar window ionization rate meter placed at the location of the right breast of an Alderson-RANDO anthropomorphic phantom was used to measure direct and scatter radiation reaching the female breast during radiotherapy of the facial region (as given for acne). The results indicate that scatter doses are very small; they are influenced by radiation quality and the use or nonuse of a treatment cone. Quantitative risk estimates show that the very small risk of breast cancer induction can be reduced even further by the use of proper radiation protection measures.

References
1.
Simon N . Breast cancer induced by radiation. Relation to mammography and treatment of acne. JAMA. 1977; 237(8):789-90. View

2.
Myrden J, Hiltz J . Breast cancer following multiple fluoroscopies during artificial pneumothorax treatment of pulmonary tuberculosis. Can Med Assoc J. 1969; 100(22):1032-4. PMC: 1946008. View

3.
Delarue N, Gale G, Ronald A . Multiple fluoroscopy of the chest: carcinogenicity for the female breast and implications for breast cancer screening programs. Can Med Assoc J. 1975; 112(12):1405-13. PMC: 1956229. View

4.
Goldschmidt H . Dermatologic radiotherapy and thyroid cancer. Arch Dermatol. 1977; 113(3):362-4. View

5.
Goldschmidt H, Sherwin W . Reactions to ionizing radiation. J Am Acad Dermatol. 1980; 3(6):551-79. DOI: 10.1016/s0190-9622(80)80067-3. View