Effects of a Reduction in Crystal Thickness on Anger-camera Performance

Overview
Journal J Nucl Med
Specialty Nuclear Medicine
Date 1979 Sep 1
PMID 536845
Citations 1
Authors
Affiliations
Soon will be listed here.
Abstract

The performances of two commerical gamma cameras, with crystals 1.3 and 0.6 cm thick, are compared. Phantom studies simulating clinical conditions showed no significant difference in performance at 140 keV. At 68 keV, the thinner crystal gave marginal improvements in camera performance with phantoms simulating clinical conditions. Frequent use with very low-energy emitters, such as Tl-201, would be needed to justify the expense required to refit the 1.3 cm camera with a 0.6 cm crystal.