» Articles » PMID: 40034211

Comparison of Swept-Source Anterior Segment Ocular Coherence Tomography and Gonioscopy in Detecting Anterior Chamber Angle Closure

Overview
Journal Clin Ophthalmol
Publisher Dove Medical Press
Specialty Ophthalmology
Date 2025 Mar 4
PMID 40034211
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To compare the diagnostic accuracy of swept-source anterior segment optical coherence tomography (AS-OCT) with gonioscopy in detecting angle closure among narrow-angle suspected eyes.

Patients And Methods: A total of 125 eyes determined narrow anterior chamber angles suspected by Van Herick's technique grade 0, 1, or 2 were recruited. AS-OCT was undergone before any contact procedure. Intraocular pressure (IOP) measurements using applanation tonometry and gonioscopy were performed, respectively. AS-OCT images were blindly interpreted by 3 glaucoma specialists, involving both qualitative assessments at 0, 90, 180, and 270 degrees and quantitative analysis using iris-trabecular contact (ITC) index. The sensitivity and specificity of AS-OCT, compared with gonioscopy-the gold standard for identifying anterior chamber angle closure-were calculated.

Results: The mean age was 61±12.9 years, and females constituted 67.6%. Eyes classified by Van Herick's technique as grade 0, 1 and 2 were 16%, 58.40%, and 25.60%, respectively. Closed-angle identified by gonioscopy and AS-OCT were 100 eyes (80%) and 102 eyes (81.60%), respectively. The sensitivity and specificity of AS-OCT in detecting angle-closure were 92.16% (95% CI, 87.44-96.87) and 73.91% (66.22-81.61), respectively. For angle-closure identified by ITC index ≥ 55%, sensitivity was 86.36% (77.53-95.20) and specificity was 92.86 (86.23-99.49). The inter-observer agreement of gonioscopy was moderate (Kappa = 0.55), while the intra-observer and inter-observer agreement of AS-OCT was substantial (Kappa = 0.71-0.80 and Kappa = 0.69, respectively).

Conclusion: AS-OCT enables a contactless qualitative and quantitative assessment of angle-closure in narrow-angle suspected eyes, demonstrating high sensitivity, acceptable specificity, and good inter-observer and intra-observer reliability.

References
1.
Ho S, Baskaran M, Zheng C, Tun T, Perera S, Narayanaswamy A . Swept source optical coherence tomography measurement of the iris-trabecular contact (ITC) index: a new parameter for angle closure. Graefes Arch Clin Exp Ophthalmol. 2012; 251(4):1205-11. DOI: 10.1007/s00417-012-2158-6. View

2.
Lavanya R, Foster P, Sakata L, Friedman D, Kashiwagi K, Wong T . Screening for narrow angles in the singapore population: evaluation of new noncontact screening methods. Ophthalmology. 2008; 115(10):1720-7, 1727.e1-2. DOI: 10.1016/j.ophtha.2008.03.015. View

3.
Sakata L, Lavanya R, Friedman D, Aung H, Gao H, Kumar R . Comparison of gonioscopy and anterior segment ocular coherence tomography in detecting angle closure in different quadrants of the anterior chamber angle. Ophthalmology. 2007; 115(5):769-74. DOI: 10.1016/j.ophtha.2007.06.030. View

4.
Nongpiur M, Aboobakar I, Baskaran M, Narayanaswamy A, Sakata L, Wu R . Association of Baseline Anterior Segment Parameters With the Development of Incident Gonioscopic Angle Closure. JAMA Ophthalmol. 2017; 135(3):252-258. DOI: 10.1001/jamaophthalmol.2016.5847. View

5.
Porporato N, Baskaran M, Tun T, Sultana R, Tan M, Quah J . Understanding diagnostic disagreement in angle closure assessment between anterior segment optical coherence tomography and gonioscopy. Br J Ophthalmol. 2019; 104(6):795-799. DOI: 10.1136/bjophthalmol-2019-314672. View