Using a Hand-Held Icterometer to Screen for Neonatal Jaundice: Validation, Feasibility, and Acceptability of the Bili-Ruler in Kumasi, Ghana

Overview
Publisher MDPI
Date 2025 Jan 25
PMID 39857549
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Neonatal jaundice (NNJ) remains a leading cause of newborn mortality in much of sub-Saharan Africa. We sought to examine the validity of using a hand-held icterometer as a screening tool to determine which newborns need further assessment. Additionally, we sought to assess the feasibility of its use among mothers.

Methods: We recruited and trained healthcare workers at one large district hospital in Ghana to use a hand-held icterometer known as the Bili-Ruler. We recruited mothers of 341 newborns aged 0 to 2 weeks at the same hospital. Mothers watched a standardized training video, after which they blanched the skin of the newborn's nose and compared it with the yellow shades numbered one to six on the icterometer. Each newborn was also assessed with a transcutaneous bilirubin meter (TCB). Research assistants and health care workers screened the same newborns, recorded their scores separately, and were blinded to each other's readings. In the second phase of this study, we recruited 100 new mothers to take the Bili-Ruler home with them, instructing them to check their newborns twice daily. We interviewed them 1-2 weeks later to determine the acceptability and feasibility of its use.

Results: Out of 341 newborns screened, 20 had elevated TCB indicative of hyperbilirubinemia. Healthcare workers' Bili-Ruler ratings had a strong and significant correlation with TCB scores, as did the ratings of researchers and mothers. When comparing Bili-Ruler scores against TCB, sensitivity across all three raters was 80% (95% CI 75.6-84.3), specificity ranged from 61.1% (healthcare providers) to 66.7% (researchers), positive predictive value ranged from 11.4% (healthcare providers) to 13.0% (researchers), and negative predictive value was 98.0% or higher across all raters. Area under the ROC curve ranged from 0.71 for healthcare providers to 0.73 for researchers. Mothers AUC was 0.72. In terms of acceptability and feasibility, the Bili-Ruler was widely accepted by the mothers and family. In total, 98% of mothers reported using it, and 90.8% used it 3 or more days in the first week after birth. Moreover, 89.8% used it more than twice per day.

Conclusions: A hand-held, low-tech icterometer is an important potential mechanism for improving early jaundice identification in low-resource settings. Further studies using larger sample sizes with a higher prevalence of hyperbilirubinemia are warranted.

References
1.
Brown A, Kim M, Wu P, Bryla D . Efficacy of phototherapy in prevention and management of neonatal hyperbilirubinemia. Pediatrics. 1985; 75(2 Pt 2):393-400. View

2.
Bhutani V, Zipursky A, Blencowe H, Khanna R, Sgro M, Ebbesen F . Neonatal hyperbilirubinemia and Rhesus disease of the newborn: incidence and impairment estimates for 2010 at regional and global levels. Pediatr Res. 2013; 74 Suppl 1:86-100. PMC: 3873706. DOI: 10.1038/pr.2013.208. View

3.
Weiner B, Lewis C, Stanick C, Powell B, Dorsey C, Clary A . Psychometric assessment of three newly developed implementation outcome measures. Implement Sci. 2017; 12(1):108. PMC: 5576104. DOI: 10.1186/s13012-017-0635-3. View

4.
Okwundu C, Olowoyeye A, Uthman O, Smith J, Wiysonge C, Bhutani V . Transcutaneous bilirubinometry versus total serum bilirubin measurement for newborns. Cochrane Database Syst Rev. 2023; 5:CD012660. PMC: 10167941. DOI: 10.1002/14651858.CD012660.pub2. View

5.
Darmstadt G, Hamer D, Carlin J, Jeena P, Mazzi E, Narang A . Validation of visual estimation of neonatal jaundice in low-income and middle-income countries: a multicentre observational cohort study. BMJ Open. 2022; 11(12):e048145. PMC: 8720979. DOI: 10.1136/bmjopen-2020-048145. View