Development and Internal Validation of an AI-Enabled Cuff-less, Non-invasive Continuous Blood Pressure Monitor Across All Classes of Hypertension

Overview
Publisher Springer
Date 2025 Jan 13
PMID 39804564
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Non-invasive, continuous blood pressure monitoring technologies require additional validation beyond standard cuff-based methods. This study evaluates a non-invasive, multiparametric wearable cuffless blood pressure (BP) diagnostic monitor across all hypertension classes with diverse subjects.

Methods: A prospective, multicenter study assessed Nanowear's SimpleSense-BP performance, including induced and natural BP changes, significant BP variations (Systolic BP (SBP) ≥ ± 15 mm Hg and Diastolic BP (DBP) ≥ ± 10 mm Hg), and reference input value validity over 4 weeks.

Results: 303 subjects (18-83 yrs; 50.16% Female) participated in algorithmic development and validation (Normal - 35%, Prehypertensive - 24%, Stage 1 - 24%, Stage 2 - 17%). 54 subjects were tested for induced change performance, 149 exhibited significant changes, and 91 validated reference value duration.

Conclusions: The study clinically validated a continuous, AI-based BP diagnostic monitor using non-invasive wearable data. Further testing on diverse populations and external validation are recommended. The protocol was inspired by ISO 81060-2 and IEEE 1708:2019 standards.

References
1.
Cohen J, Lotito M, Trivedi U, Denker M, Cohen D, Townsend R . Cardiovascular Events and Mortality in White Coat Hypertension: A Systematic Review and Meta-analysis. Ann Intern Med. 2019; 170(12):853-862. PMC: 6736754. DOI: 10.7326/M19-0223. View

2.
Mukkamala R, Shroff S, Landry C, Kyriakoulis K, Avolio A, Stergiou G . The Microsoft Research Aurora Project: Important Findings on Cuffless Blood Pressure Measurement. Hypertension. 2022; 80(3):534-540. PMC: 9931644. DOI: 10.1161/HYPERTENSIONAHA.122.20410. View

3.
Shyam Kumar P, Rai P, Ramasamy M, Varadan V, Varadan V . Multiparametric cloth-based wearable, SimpleSense, estimates blood pressure. Sci Rep. 2022; 12(1):13059. PMC: 9338290. DOI: 10.1038/s41598-022-17223-x. View

4.
Bangalore S, Gong Y, Cooper-DeHoff R, Pepine C, Messerli F . 2014 Eighth Joint National Committee panel recommendation for blood pressure targets revisited: results from the INVEST study. J Am Coll Cardiol. 2014; 64(8):784-93. PMC: 4193384. DOI: 10.1016/j.jacc.2014.05.044. View

5.
Mukkamala R, Yavarimanesh M, Natarajan K, Hahn J, Kyriakoulis K, Avolio A . Evaluation of the Accuracy of Cuffless Blood Pressure Measurement Devices: Challenges and Proposals. Hypertension. 2021; 78(5):1161-1167. PMC: 8516718. DOI: 10.1161/HYPERTENSIONAHA.121.17747. View