Introduction: Opening the incubator side wall to insert a non-sterile length-measuring device carries the risk of microbial contamination and thermal instability for preterm infants. To reduce this inconvenience, a laser-based length-measuring board is proposed to measure body length from outside the incubator.

Methods: This device has two laser-line-shaped cursors which can be pointed to opposite ends of a segment to be measured. It is attached to the outer side of one of the incubator's side walls in such a manner as to ensure that its axis is parallel to the longitudinal axis of the segment. To validate the measurements made with this model, a calibrated caliper consisting of a conventional rigid length-measuring board with a resolution of 0.05 mm was constructed to serve as a reference. Crown-heel length was measured in a sample of 45 infants, including 32 preterm and 13 term infants of corrected gestational age at the time of measurement.

Results: Good intra-observer variability was obtained. Near-perfect statistical agreement was found between measurements with both devices, with concordance correlation coefficients of 0.994 (95% CI: 0.990; 0.996) in preterm infants and 0.994 (95% CI: 0.988, 0.998) in infants at term. The clinical relevance of the agreement between measurements was assessed by a Bland-Altman plot, and the difference may reach clinical relevance (up to 1 cm) but without evidence of proportional bias.

Conclusion: The proposed validated laser-based length-measuring board offers a suitable alternative to conventional length-measuring boards for contactless measurement of infant body length.

References
1.
Brennan A, Murphy B, Kiely M . Optimising preterm nutrition: present and future. Proc Nutr Soc. 2016; 75(2):154-61. DOI: 10.1017/S0029665116000136. View

2.
Shinwell E, Shlomo M . Measured length of normal term infants changes over the first two days of life. J Pediatr Endocrinol Metab. 2003; 16(4):537-40. DOI: 10.1515/jpem.2003.16.4.537. View

3.
Pereira-da-Silva L, Virella D, Fusch C . Nutritional Assessment in Preterm Infants: A Practical Approach in the NICU. Nutrients. 2019; 11(9). PMC: 6770216. DOI: 10.3390/nu11091999. View

4.
Carrasco J, Jover L . Estimating the generalized concordance correlation coefficient through variance components. Biometrics. 2004; 59(4):849-58. DOI: 10.1111/j.0006-341x.2003.00099.x. View

5.
Ismail L, Puglia F, Ohuma E, Ash S, Bishop D, Carew R . Precision of recumbent crown-heel length when using an infantometer. BMC Pediatr. 2016; 16(1):186. PMC: 5109659. DOI: 10.1186/s12887-016-0725-4. View