Quantification Approaches in Non-Target LC/ESI/HRMS Analysis: An Interlaboratory Comparison

Abstract

Nontargeted screening (NTS) utilizing liquid chromatography electrospray ionization high-resolution mass spectrometry (LC/ESI/HRMS) is increasingly used to identify environmental contaminants. Major differences in the ionization efficiency of compounds in ESI/HRMS result in widely varying responses and complicate quantitative analysis. Despite an increasing number of methods for quantification without authentic standards in NTS, the approaches are evaluated on limited and diverse data sets with varying chemical coverage collected on different instruments, complicating an unbiased comparison. In this interlaboratory comparison, organized by the NORMAN Network, we evaluated the accuracy and performance variability of five quantification approaches across 41 NTS methods from 37 laboratories. Three approaches are based on surrogate standard quantification (parent-transformation product, structurally similar or close eluting) and two on predicted ionization efficiencies (RandFor- and MLR-). Shortly, HPLC grade water, tap water, and surface water spiked with 45 compounds at 2 concentration levels were analyzed together with 41 calibrants at 6 known concentrations by the laboratories using in-house NTS workflows. The accuracy of the approaches was evaluated by comparing the estimated and spiked concentrations across quantification approaches, instrumentation, and laboratories. The RandFor- approach performed best with a reported mean prediction error of 15× and over 83% of compounds quantified within 10× error. Despite different instrumentation and workflows, the performance was stable across laboratories and did not depend on the complexity of water matrices.

Citing Articles

Examining environmental matrix effects on quantitative non-targeted analysis estimates of per- and polyfluoroalkyl substances.

Pu S, McCord J, Dickman R, Sayresmith N, Sepman H, Kruve A Anal Bioanal Chem. 2025; .

PMID: 40014069 DOI: 10.1007/s00216-025-05796-1.


Unravel the in-Source Fragmentation Patterns of Per- and Polyfluoroalkyl Substances during Analysis by LC-ESI-HRMS.

Wang K, Wang R, Shan W, Yang Z, Chen Y, Wang L Environ Sci Technol. 2024; 58(51):22766-22776.

PMID: 39668558 PMC: 11674715. DOI: 10.1021/acs.est.4c08442.


A Semi-Quantitative Approach to Nontarget Compositional Analysis of Complex Samples.

Evans R, Bryant D, Voliotis A, Hu D, Wu H, Syafira S Anal Chem. 2024; 96(46):18349-18358.

PMID: 39508740 PMC: 11579983. DOI: 10.1021/acs.analchem.4c00819.

References
1.
Souihi A, Mohai M, Palm E, Malm L, Kruve A . MultiConditionRT: Predicting liquid chromatography retention time for emerging contaminants for a wide range of eluent compositions and stationary phases. J Chromatogr A. 2022; 1666:462867. DOI: 10.1016/j.chroma.2022.462867. View

2.
Richardson S, Ternes T . Water Analysis: Emerging Contaminants and Current Issues. Anal Chem. 2017; 90(1):398-428. DOI: 10.1021/acs.analchem.7b04577. View

3.
Tisler S, Kilpinen K, Pattison D, Tomasi G, Christensen J . Quantitative Nontarget Analysis of CECs in Environmental Samples Can Be Improved by Considering All Mass Adducts. Anal Chem. 2023; 96(1):229-237. PMC: 10782417. DOI: 10.1021/acs.analchem.3c03791. View

4.
Escher B, Fenner K . Recent advances in environmental risk assessment of transformation products. Environ Sci Technol. 2011; 45(9):3835-47. DOI: 10.1021/es1030799. View

5.
Kalogiouri N, Aalizadeh R, Thomaidis N . Investigating the organic and conventional production type of olive oil with target and suspect screening by LC-QTOF-MS, a novel semi-quantification method using chemical similarity and advanced chemometrics. Anal Bioanal Chem. 2017; 409(23):5413-5426. DOI: 10.1007/s00216-017-0395-6. View