Diagnostic Role of Sysmex Hematology Analyzer in the Detection of Malaria: A Systematic Review and Meta-analysis

Overview
Journal PLoS One
Date 2024 Sep 6
PMID 39240990
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Malaria control depends primarily on rapid and accurate diagnosis followed by successful treatment. Light microscopy is still used as a gold standard method for the diagnosis of malaria. The Sysmex hematology analyzer is a novel method for malaria detection. Therefore, the aim of this review was to investigate the diagnostic accuracy of the Sysmex hematology analyzer for malaria diagnosis.

Methods: Electronic databases like PubMed, PubMed Central, Science Direct databases, Google Scholar, and Scopus were used to find relevant articles from April to June 14, 2023. The studies' methodological quality was assessed using the Quality Assessment of Diagnostic Accuracy Studies-2 tool. Using Review Manager 5.4.1, the estimates of sensitivity and specificity, as well as their 95% confidence intervals, were shown in forest plots. Midas software in Stata 14.0 was utilized to calculate the summary estimates of sensitivity, specificity, positive likelihood ratio, negative likelihood ratio, and diagnostic odds ratio. Heterogeneity was assessed by using I2 statistics. In addition, publication bias was assessed using a funnel plot and Deeks' test. Sub-group and meta- regression analysis were also performed.

Results: A total of 15 studies were assessed for diagnostic accuracy. The sensitivity and specificity of Sysmex hematology analyzer for studies ranged from 46% to 100% and 81% to 100%, respectively. The summary estimate of sensitivity and specificity of Sysmex hematology analyzer were 95% (95% CI: 85%-99%) and 99% (95% CI: 97%-100%), respectively. It had excellent diagnostic accuracy. There were significant heterogeneity among the studies included in this meta-analysis. The summary estimate of sensitivity and specificity of Sysmex hematology analyzer using polymerase chain reaction as the gold standard was 97.6% (95% CI: 83.2, 99.7) and 99.4% (98.5, 99.8), respectively.

Conclusion: In this review, Sysmex hematology analyzer had excellent diagnostic accuracy. Therefore, it could be used as an alternate diagnostic tool for malaria diagnosis in the hospital and health center.

Trial Registration: Systematic review registration PROSPERO (2023: CRD42023427713). https://www.crd.york.ac.uk/prospero/display_record.php?ID=CRD42023427713.

References
1.
Ochola L, Vounatsou P, Smith T, Mabaso M, Newton C . The reliability of diagnostic techniques in the diagnosis and management of malaria in the absence of a gold standard. Lancet Infect Dis. 2006; 6(9):582-8. DOI: 10.1016/S1473-3099(06)70579-5. View

2.
Hofmann N, Gruenberg M, Nate E, Ura A, Rodriguez-Rodriguez D, Salib M . Assessment of ultra-sensitive malaria diagnosis versus standard molecular diagnostics for malaria elimination: an in-depth molecular community cross-sectional study. Lancet Infect Dis. 2018; 18(10):1108-1116. DOI: 10.1016/S1473-3099(18)30411-0. View

3.
Komaki-Yasuda K, Kutsuna S, Kawaguchi M, Kamei M, Uchihashi K, Nakamura K . Clinical performance testing of the automated haematology analyzer XN-31 prototype using whole blood samples from patients with imported malaria in Japan. Malar J. 2022; 21(1):229. PMC: 9338637. DOI: 10.1186/s12936-022-04247-x. View

4.
Zuluaga-Idarraga L, Rios A, Sierra-Cifuentes V, Garzon E, Tobon-Castano A, Takehara I . Performance of the hematology analyzer XN-31 prototype in the detection of Plasmodium infections in an endemic region of Colombia. Sci Rep. 2021; 11(1):5268. PMC: 7933134. DOI: 10.1038/s41598-021-84594-y. View

5.
Yoo J, Song J, Lee K, Sun Y, Kim Y, Park T . Automated detection of malaria-associated pseudoeosinophilia and abnormal WBC scattergram by the Sysmex XE-2100 hematology analyzer: a clinical study with 1,801 patients and real-time quantitative PCR analysis in vivax malaria-endemic area. Am J Trop Med Hyg. 2010; 82(3):412-4. PMC: 2829901. DOI: 10.4269/ajtmh.2010.09-0560. View