» Articles » PMID: 39157062

A Heating and Cooling Stage With Fast Temporal Control for Biological Applications

Overview
Publisher IEEE
Date 2024 Aug 19
PMID 39157062
Authors
Affiliations
Soon will be listed here.
Abstract

The study of biological processes involving live microscopy techniques requires adequate temperature control to respect the physiology of the organism under study. We present here a design strategy for a microscope temperature stage based on thermoelectric elements. The design allows the user to access a range of temperatures below and above room temperature and can accommodate samples of different geometries. In addition, by cooling simultaneously the sample insert and the objective, we minimize the temperature gradients along the sample for large magnification objectives requiring immersion oil. We illustrate how this design can be used to study the physiology of the zebrafish embryo over the temperature tolerance of this species. We envision that this device could benefit the communities using model and non-model organisms with physiological temperatures different from typical mammalian cell culture incubation in biomedical research.

References
1.
Schweizer M, Dieterich A, Triebskorn R, Kohler H . Drifting Away of a FET Endpoint: The Heart Rate in Danio rerio Embryos is Extremely Sensitive to Variation in Ambient Temperature. Bull Environ Contam Toxicol. 2017; 99(6):684-689. DOI: 10.1007/s00128-017-2196-1. View

2.
Maia Chagas A, Prieto-Godino L, Arrenberg A, Baden T . The €100 lab: A 3D-printable open-source platform for fluorescence microscopy, optogenetics, and accurate temperature control during behaviour of zebrafish, Drosophila, and Caenorhabditis elegans. PLoS Biol. 2017; 15(7):e2002702. PMC: 5515398. DOI: 10.1371/journal.pbio.2002702. View

3.
Rayani K, Lin E, Craig C, Lamothe M, Shafaattalab S, Gunawan M . Zebrafish as a model of mammalian cardiac function: Optically mapping the interplay of temperature and rate on voltage and calcium dynamics. Prog Biophys Mol Biol. 2018; 138:69-90. DOI: 10.1016/j.pbiomolbio.2018.07.006. View

4.
Casquillas G, Fu C, Le Berre M, Cramer J, Meance S, Plecis A . Fast microfluidic temperature control for high resolution live cell imaging. Lab Chip. 2010; 11(3):484-9. DOI: 10.1039/c0lc00222d. View

5.
Slangen P, Porat Y, Mertz M, van den Broek B, Jalink K, de Gooijer M . Protocol for live-cell imaging during Tumor Treating Fields treatment with Inovitro Live. STAR Protoc. 2022; 3(2):101246. PMC: 8971986. DOI: 10.1016/j.xpro.2022.101246. View